English

Xi'an Yunyi Instrument Co., Ltd
Home>Products>GXY electric self weight micro pressure gauge
GXY electric self weight micro pressure gauge
Introduction to Self Weight Pressure Gauge: The DXY-1 electric self weight pressure gauge adopts the self weight principle and can provide high-precis
Product details
Introduction to Self Weight Pressure Gauge:

The DXY-1 electric self weight pressure gauge adopts the self weight principle and can provide high-precision standard pressure signals. It is one of the essential instruments for pressure calibration and high-precision pressure reference. The electric pressure gauge can calibrate pressure transmitters, pressure gauges, pressure switches and other pressure instruments, and can also provide standard pressure signals for scientific research institutions. At present, the DXY series electric pressure gauge fills the gap of gas pistons in China.
Principle of self weight pressure gauge:
The DXY-1 electric self weight pressure gauge is equipped with dual pressure reducing valves and a constant current controller to ensure constant gas source pressure, flow, and pressure. Under constant flow and pressure conditions, the piston system is suspended above a constant gas pressure, and excess gas will overflow from the overflow hole to ensure that the pressure generated by the piston system is equal to the gas pressure. The gas pressure is output through a stabilizing device, and the gas source pressure is balanced and converted into a standard pressure signal.

Advantages of Self Weight Pressure Gauge:

1. Built in micro compressor, no additional air source required
2. Self balancing principle, stable without manual pressurization
3. Fast calibration speed, faster than traditional calibration devices based on digital meters
4. Equipped with a voltage and current measurement module, it can measure the feedback signal of the transmitter and output 24V
5. The pressure standard is based on pure mechanical principles, with good accuracy and stability over the years.
6. Easy to operate, no additional professional knowledge required for operation
7. Accuracy refers to the reading accuracy, which is higher than the precision of digital meter calibration
8. Two pressure transmitters can be calibrated simultaneously

Technical parameters of self weight pressure gauge:

Pressure range: 1-300KPa (minimum pressure is 1KPa)
Accuracy: 0.05% (reading accuracy)
Flow rate: 0.5L/min
Current: 4-20mA 0-10V 0.05%
Load output: 24V (supplying power to the transmitter)
Power supply: 220V
Power consumption: 30W

The use of self weight pressure gauges:
  1. Install the tested pressure transmitter
  2. Close the pressure relief valve
  3. Connect the probe to the positive and negative terminals of the transmitter output.
  4. Turn on the power
  5. Place weights
  6. Stable pressure within 5 seconds
  7. The gravity generated by the weight is the current pressure
  8. Check the output current of the transmitter and compare whether it is accurate
Parameters and Model:
DXY-1 1-150KPa
DXY-2 1-250KPa
DXY-3 1-400KPa
Online inquiry