High purity gas analysis gas chromatographG51-GH
High purity gas analysis gas chromatographsummary:
G51 is an economical high-purity gas analyzer suitable for impurity analysis in gases such as hydrogen, oxygen, argon, nitrogen, helium, neon, krypton, xenon, methane, carbon dioxide, etc; Detect the minimum concentrations of H2, O2, and N2 above 1:1ppm, and CO, CH4, and CO2 above 2.01ppm, meeting the requirements for high-purity gas analysis with purity ranging from 99.99% to 99.999%.
If ppb ppt level impurities need to be detected, helium ionization detectors PHHID and plasma discharge detectors EPD with central cutting systems can be selected, with a minimum concentration of up to 100ppm for detection
system configuration:(For reference only)
(1)G51Gas chromatograph host (wide range)FIDdetector+high sensitivityTCDDetector)
(2)Customized valve system
(3)Chromatographic column system
(4)Modular methane converter
(6)Mini pillar box
(7)Standard gas4Lor8L
(8)Stainless steel sampling valve
(9)Carrier gas (purity)>99.999%According to the testing components
(10)Special stainless steel pressure reducing valve for carrier gas
(11)High purity hydrogen generator
(12)Low noise air pump
(13)309Series network version anti control chromatography workstation
Classic spectrumExample diagram:Analysis of Impurities in High Purity Tungsten Hexafluoride
Standard gas component content: 5ppm
Injection method: valve injection
Detector: HTCD
Peak order:1air2、CF4;3、CO2;4、SF6
Main features and performance indicators of G51Pro/G51ProA
17Inch color touch screen display, the host can provide RS232 communication port and LANE network port.
2. Manual adjustment of gas path and fully electronic gas pathEPCOptional control, networked communication, computer online full control operation.
3. All of themThe detector and injection port can both be equipped with electronic pressure monitoring mode.
aThere are 9 or 13 electronic pressure measurement channels.
bPressure sensor:
l Accuracy:<± 2% of full scale
l Reproducibility < ± 0.05Kpa
l Temperature coefficient < ± 0.01Kpa/℃
l Range: 0-0.3Kpa or 0-0.6Kpa
4. All detectorsBoth injection ports can be equipped with high-precision electronic gas circuitsAPC/AFCController, which can set the gas flow rate, purge flow rate, and split ratio of the chromatographic column in the display screen and software. It can also set the flow rates of hydrogen, air, and tail gas, making it easy and efficient to operate;
a、 The EPC electronic pneumatic control accuracy is 0.001psi, ensuring better retention time reproducibility and more consistent and reliable results.
b、 EPC working mode: constant current, constant voltage, and split current mode.
cProgram pressure control: 4th order
d、 EPC working gas: N2、H2、Air、He、Ar
e、 EPC control range: pressure 0-150psi, flow rate 0-500sccm or 0-1000sccm (optional)
f、 EPC control accuracy: pressure 0.001psi; Flow rate of 0.01 sccm
gSplit ratio setting: 0-2000:1 or 0-10000:1 (customized)
hSplit flow rate setting: 0-200 sccm or 0-1000 sccm
iUp to 18 APC/AFC installations are available, suitable for diversified and multitasking analysis.
5. The temperature can be raised to 450 ℃, suitable for the analysis of high boiling point samples.
6. Supports dual column box dual rear door opening mode and liquid nitrogen mode.
7It has 8 external event controls and 3 external trigger signal inputs.
8. Optional gas cut-off protection function. When the carrier gas pressure is below 0.1MPa, the instrument's main control circuit will automatically shut down and enter standby mode to protect the detector and chromatographic column.