★ Overview
Differential capacitive sensors are currently the most mature technology for measuring pressure and differential pressure. The FB1151 pressure transmitter draws on advanced international technology and, while maintaining and promoting the excellent performance of similar foreign products, adopts an exquisite external structure with independent intellectual property rights, special processes and equipment, and advanced electronic circuit design. The FB1151 pressure transmitter is an ideal upgraded product for traditional 1151 pressure transmitters.
★ Scope of Application
The FB1151 pressure transmitter is suitable for accurately measuring micro differential pressure to large differential pressure, low pressure to high pressure, liquid level, vacuum degree, and specific gravity; Combined with a throttling device, flow can also be measured.
★ Features
Exquisite exterior design;
High precision, diverse varieties, wide measurement range, and good stability (see variety specification table);
※ Adopting a two-wire working mode, optional with HART fieldbus protocol;
Using solid circuit components, the printed circuit board has a plug-in method;
Unique structural design, sturdy and strong seismic resistance;
The range and zero position can be continuously adjusted externally;
Positive migration can reach 500%, and negative migration can reach 600%;
※ Adjustable damping;
※ There are a variety of options available;
★ Technical parameters
|
● Target audience |
Liquid, gas or vapor |
|
|
● Measurement range |
gauge pressure |
0-0.13 kPa to 0-41370 kPa |
|
differential pressure |
0-0.125 kPa to 0-6890 kPa |
|
|
absolute pressure |
0-6.8 kPa to 0-6890 kPa |
|
|
● Output |
4~20mADC |
|
|
● Power supply |
Ordinary type: 12-45VDC; Explosion proof type: 24VDC; Intrinsic safety type: powered by safety barriers |
|
|
Positive and negative transfer |
Regardless of the output, after positive and negative migration, the upper and lower limits of the range must not exceed the limit of the range Maximum migration amount: 500% of the minimum calibration range Maximum negative transfer: 600% of the minimum calibration range |
|
|
● Temperature range |
The working temperature of conventional transmitters is -25 ℃ to+70 ℃; Equipped with on-site indicator table -15 ℃~+70 ℃ Storage temperature: -40 ℃~+70 ℃ Amplifier operating temperature range (ambient temperature): -29 ℃~+93 ℃ (LT type: -30 ℃~+70 ℃) Working temperature of silicone oil filled measuring element (LT type medium temperature): -40 ℃ to+149 ℃ When filling high-temperature silicone oil into remote transmitters: -15~+315 ℃, ordinary silicone oil: -40~+149 ℃ |
|
|
● Humidity range |
Relative humidity is 1-100% |
|
|
● Sensor volume change |
Less than 0.16cm3 |
|
|
● Indicator table |
(On site) pointer type; digital display |
|
|
Explosion proof type |
Explosion proof ExdIICT6; Intrinsic safety ExiaICT6 |
|
|
● Range and zero point |
External continuous adjustable |
|
|
● Overpressure limit |
Application range below 6MPa: 0 (absolute pressure) to 14MPa; Application range of 20MPa: 0 (absolute pressure) to 32MPa; Application range of 42MPa: 0 (absolute pressure) to 52MPa; The transmitter is undamaged, and the flange can withstand 60MPa |
|
|
● Damping |
Continuously adjustable between 0.2 and 16 seconds |
|
|
● Start time |
2 seconds, no preheating required |
|
|
● Vibration impact |
In any axial direction, with a frequency of 200Hz, the error caused is less than ± 0.05% of the maximum range, and the range code 2 (micro differential pressure) is ± 0.25% |
|
|
● Power supply impact |
Less than 0.005%/V of the output range |
|
|
● Load impact |
When the power supply is stable, there is no load impact |
|
|
● Installation location impact |
The maximum zero point error that can be generated is 0.24 kPa, but it can be calibrated and has no impact on the measurement range |
|
|
● Accuracy |
See the table below |
|
|
Model and name |
measurement range |
precision |
stability |
|
1151DP Low, medium, and high differential pressure transmitter |
0-1.24 kPa to 186.45 kPa (Range codes 3, 4, 5) |
±0.2% |
<0.1% FS/year |
|
1151DP Large differential pressure transmitter |
0-117.21 kPa to 6890 kPa (Range codes 6, 7, 8) |
±0.25% |
<0.1% FS/year |
|
1151HP High static pressure differential pressure transmitter |
0-6.22 kPa to 2070 kPa (Range codes 4, 5, 6, 7) |
±0.25% |
<0.1% FS/year |
|
1151GP Pressure transmitter |
0-1.24 kPa to 41370kPa (Range code 3-0) |
±0.25% |
<0.1% FS/year |
|
1151GP/DP Remote differential pressure transmitter |
0-6.22 kPa to 6890 kPa (Range codes 4, 5, 6, 7, 8) |
±0.5% |
<0.25% FS/year |
|
1151AP Absolute pressure transmitter |
0-6.77 kPa to 6890 kPa (Range codes 4, 5, 6, 7, 8) |
±0.25% |
<0.1FS%/year |
|
1151DR Micro differential pressure transmitter |
0-0.12 kPa to 1.49 kPa (Range code 2) |
±0.5% |
<0.25% FS/year |
|
● Temperature impact At the maximum range, Change every 28 ℃ |
Range code Or name |
Accuracy class |
||
|
Level 0.2 |
0.25 level |
Level 0.5 |
||
|
2 |
/ |
/ |
≤±0.54%/ |
|
|
3 |
≤±0.20% |
≤±0.30% |
≤±0.45% |
|
|
4. 5, 6, 7, 8 |
≤±0.20% |
≤±0.20% |
≤±0.30% |
|
|
Remote transmission, liquid level |
/ |
/ |
≤±0.60% |
|
|
●Static pressure influence (At maximum range) |
Range code Or name |
Accuracy class |
||
|
Level 0.2 |
0.25 level |
Level 0.5 |
||
|
2 |
/ |
/ |
≤±1% |
|
|
3 |
≤±0.2 |
≤±0.5% |
≤±0.6% |
|
|
4. 5, 6, 7, 8 |
≤±0.20% |
≤±0.30% |
≤±0.6% |
|
|
4. 5, 6, 7 (high static pressure) |
/ |
≤±2% |
≤±2.5% |
|
|
※performance metrics |
|
|
● Structural materials |
Isolation membrane: 316 stainless steel, Hastelloy C, Monel alloy, Tantalum O-ring (medium contact area): Fluororubber Flange: 316 stainless steel Filling fluid: silicone oil or inert oil Electrical casing: polyurethane baked paint low copper cast aluminum alloy |
|
● Pressure connector |
The connecting screw hole on the pressure chamber is 1/4NPT, and the connecting screw hole on the impulse joint is 1/2NPT. The center distance is shown in the external dimension drawings of various models The signal cable connection screw hole is M20 × 1.5 |
|
● Shell protection level |
IP65 |
|
● Weight |
4.2kg (excluding options, except for flanged transmitters)
|
«FB1151APType Absolute Pressure Transmitter Selection Table
|
FB1151AP |
Absolute pressure transmitter |
||||||||
|
┆ |
code name |
Measurement range (kPa) |
|||||||
|
┆ |
4 |
0-6.22~37.29 |
|||||||
|
┆ |
5 |
0-31.08~186.45 |
|||||||
|
┆ |
6 |
0-117.21~689.48 |
|||||||
|
┆ |
7 |
0-345~2068 |
|||||||
|
┆ |
8 |
0-1170~6890 |
|||||||
|
┆ |
┆ |
code name |
output |
||||||
|
┆ |
┆ |
E |
4~20mADC |
||||||
|
┆ |
┆ |
S |
Intelligent 4-20mADC two-wire system+fieldbus HART protocol |
||||||
|
┆ |
┆ |
┆ |
code name |
Structural materials |
|||||
|
┆ |
┆ |
┆ |
flange joint |
Exhaust/Drain Valve |
Isolation membrane |
Fill with liquid |
|||
|
┆ |
┆ |
┆ |
22 |
316 stainless steel |
316 stainless steel |
316 stainless steel |
silicone oil |
||
|
┆ |
┆ |
┆ |
23 |
316 stainless steel |
316 stainless steel |
Hastelloy C |
|||
|
┆ |
┆ |
┆ |
24 |
316 stainless steel |
316 stainless steel |
Monel |
|||
|
┆ |
┆ |
┆ |
25 |
316 stainless steel |
316 stainless steel |
Tantalum |
|||
|
┆ |
┆ |
┆ |
33 |
Hastelloy C |
Hastelloy C |
Hastelloy C |
|||
|
┆ |
┆ |
┆ |
44 |
Monel |
Monel |
Monel |
|||
|
┆ |
┆ |
┆ |
┆ |
code name |
Options |
||||
|
┆ |
┆ |
┆ |
┆ |
M1 |
0-100% linear indicator head |
||||
|
┆ |
┆ |
┆ |
┆ |
M3 |
3 1/2LCD digital display header |
||||
|
┆ |
┆ |
┆ |
┆ |
B1 |
Pipe bending bracket |
||||
|
┆ |
┆ |
┆ |
┆ |
B2 |
Board mounted bending bracket |
||||
|
┆ |
┆ |
┆ |
┆ |
B3 |
Pipe mounted flat bracket |
||||
|
┆ |
┆ |
┆ |
┆ |
D0 |
The relief valve is located at the rear of the pressure chamber |
||||
|
┆ |
┆ |
┆ |
┆ |
D1 |
The side relief valve is located at the upper part of the pressure chamber |
||||
|
┆ |
┆ |
┆ |
┆ |
D2 |
The side relief valve is located at the lower part of the pressure chamber |
||||
|
┆ |
┆ |
┆ |
┆ |
C1 |
1/2NPT pressure transition joint rear welded with a φ 14 pressure pipe |
||||
|
┆ |
┆ |
┆ |
┆ |
C2 |
T-shaped threaded joint M20 × 1.5 |
||||
|
┆ |
┆ |
┆ |
┆ |
C3 |
1/2NPT taper pipe female threaded joint |
||||
|
┆ |
┆ |
┆ |
┆ |
d |
Explosion proof ExdIICT6 |
||||
|
┆ |
┆ |
┆ |
┆ |
i |
Intrinsically safe ExiaICT6 |
||||
|
FB1151AP |
8 |
E |
22 |
M1 |
(Factory range) |
Example of transmitter selection |
|||
Note: When placing an order, the factory range must be specified. If not specified, the highest rated range shall be used for factory delivery.