The F-1 fluoride ion concentration meter is an electrochemical analysis instrument that uses direct potentiometry to accurately measure the fluoride ion concentration in a solution. The instrument can not only measure the concentration of fluoride ions in aqueous solutions, but also serve as a rapid measure for the chemical defense of troops against phosphorus containing toxins, sarin, and soman.
Technical indicators:
*Measurement range: mV:-499 mV~+499mV
C: 1 × 10-3~1×10-6mol/l
*Relative error: mV: 0.2% F · S ± 1 word
C: 1 × 10-3~5×10-6mol/l <5%
5 × 10-6~1×10-6mol/l <10%
*Resolution: 1mV
*Calibration method: three-point calibration
*Input impedance:> 3×1011Ω
*Continuous working hours: 10h
*Dimensions: 175 × 135 × 60 (mm)
*Weight: 0.5kg
*Power consumption:< 100 mW
characteristic:
*High integration, low power consumption
*Small size and light weight
*Composite electrode
*Concentration direct reading
*Easy to operate
*Multiple fault diagnosis functions
Applied to environmental monitoring, health and epidemic prevention, research institutes, mines and other departments, especially suitable for field testing