Introduction:
There are many low-frequency field strengths present in our living environment, and any place where electricity is generated, transmitted, and used will generate low-frequency electric fields, such as homes, workplaces, medical equipment, and even outdoor high-voltage transmission lines will generate electric fields. Wherever there is high current, there is a magnetic field; wherever there is high voltage, there is an electric field. It is almost impossible to completely avoid the influence of electric fields. It is important to understand their impact on us and reduce it through certain meanselectromagnetic radiationThe impact.
Applicable standards:
GBZ/T 189.3-2018 "Measurement of Physical Factors in the Workplace";
GB 8702-2014 Electromagnetic Environment Control Limits;
EN V 50166, EN 50366, BImschV, DIN VDE 0848, BGV B11, etc.
The relevant measuring instruments corresponding to these specifications should consider the following aspects:
- In order to obtain and establish the true response of the human body in electromagnetic fields, monitoring instruments should have isotropic and real-time functions, and must be able to simultaneously measure the values of electric and magnetic fields, as electromagnetic fields act on the human body simultaneously.
- The measuring coil must be small enough to accurately transmit measurement results even in close proximity to the magnetic field source, minimizing introduced errors.
- Because various fields can generate fluctuations, and many limits depend on time accumulation. So, it is meaningful to record for 24 hours without a computer. Therefore, measuring instruments must have sufficient continuous measurement time and large storage space.
Low frequency field with frequent occurrences:
- Power station and transformer
- High voltage line
- Electric devices and railways
- Welding, tempering, and melting industrial equipment
- Galvanizing equipment and chlorination production
- Medical equipment, MRT
Application:
1. Environmental Protection Bureau
2. Environmental monitoring unit
3. Occupational health monitoring(Center for Disease Control and PreventionEnvironmental Impact Assessment Center, Occupational Health Assessment Center, Work/Living Environment Monitoring)
4. Third party organizations (environmental monitoring units, environmental supervision companies, environmental certification agencies, etc.)
5. University research institutions
6. Enterprise R&D
7. Radio monitoring
Typical users:
AEG Electric, Deutsche Bahn, Bureau Veritas, Barrow Cumbria, Federal Office for Radiation Protection, Continental, Daimler, Henkel, DEKRA, Bosch, Bayer, Rolls Royce Renault, Schneider Electric, VDE, T Ü V, Enel, Alstom, Siemens, RWE Rheinland, E-ON, EADS Airbus, Osram, Philips, Lujun, Zhejiang Academy of Medical Sciences, Guangdong Entry Exit
characteristic
- Synchronous isotropic measurement of electric and magnetic fields
- Frequency range:5Hz – 400kHz
- Measurement range: Magnetic field 1nT-20mT, electric field 0.1V/m-100kV/m
- Handheld simple and intuitive operation can avoid measurement errors
- ± 5% high precision and ultra long time stability
- Compliant with EN V 50166, DIN VDE 0848 standard
- Can record for more than 24 hours without the need for a host or computer
- Can store 1800 records
- Four optional filters
- Switchable for one-dimensional measurement
- Can set signal tone
- Background light display
- 45 Continuous operation for hours
- Moisture and dust resistant, compliant with IP65
- German R&D and manufacturing
Special Features:Transformer working load correction can be made according to corresponding standards to make measurement data more accurate and reliable.
Software
Graph ESM-100 software: After connecting to a computer, ESM-100 can be remotely controlled through Graph ESM-100 software and record all measurement results. This software is specially designed for ESM-100 and meets EMC monitoring requirements. With just a click of the mouse, you can draw, table, and create three-dimensional spatial distribution maps. The scale and corresponding axial letters are automatically generated. Simply select the corresponding control key to directly track the measured value or display the real-time measurement range curve.
Software Features
·Visual display of measurement value chart, real-time measurement curve online display, simple operation, intuitive desktop table display, each measurement value with annotation space for setting limits
·ESM-100 memory block download burst control record
·Snapshot function
·The help function of ESM-100 remote control for magnetic field workload correction, which can output graphics and forms *
·Compatible with Windows operating system
Technical parameters:
|
magnetic field |
electric field |
measuring range |
1nT-20mT (>100KHz up to 20uT) |
0.1V/m-100KV/m (>100KHz up to 2KV/m) |
Display Range |
0nT - 20mT |
0.0V/m-100KV/m(Filter 50/16.7) |
|
10nT – 20mT |
1V/m-100KV/m(Filter HIGH LOW) |
|
15nT – 20mT |
1.5V/m-100KV/m(Filter all) |
accuracy |
1nT |
0.1V/m |
data processing |
Simultaneously processing information from six channels of three-dimensional electric and magnetic fields |
|
Optional frequency range |
5Hz – 400KHz (-3dB Limit) |
|
|
Filter all 5Hz – 400KHz |
|
|
Filter high 2KHz– 400KHz |
|
|
Filter low 5Hz – 2KHz |
|
|
Filter 50 50Hz band-pass 12db |
|
|
Filter 16 16.7Hz band-pass 12db |
|
Accuracy |
±5%(sine; 50nT-20mT;5V/m-100KV/m) ±5 Digits |
|
measurement error |
5Hz-400KHz: less than 0.3dB, with measurement accuracy of high-level professional measuring instruments |
|
Measuring speed |
Display 2Hz in decimal system |
Bar display 10Hz with 3sec peak hold |
principle |
Magnetic field: coil, isotropic |
Electric field: polar plate, isotropic |
|
The true RMS value |
|
Unidirectional positioning |
Not only can it measure in all directions in three dimensions, but it can also locate the radiation source in one direction in one dimension |
|
Electric and magnetic fields |
H=√Hx+Hy+Hz |
E=√Ex+Ey+Ey |
|
Unit: nT, µ T, mT, Automatic switching |
Unit V/m, KV/m, Automatic switching |
function |
Minimum value, maximum value, storage, one-dimensional measurement, control charging, and battery fault display |
|
record |
1800 records (6 channels, adjustable functions) |
|
time interval |
1h 8h 24 hours 48 hours 168h, "snapshot" |
|
time interval |
16 48 96 336 seconds, "snapshot" and "curve" |
|
Analog output port |
4 x 0-600 mVrms (input impedance>100kOhm) Hx Ex |
|
Battery |
Nickel metal hydride battery. (Ni-MH 6V/1.5 Ah) |
|
Usage time |
More than 45 hours, fast charging for 4 hours |
|
Host operation |
Automatically switch between host operation, fast charging, and maintenance charging modes |
|
Connect computer |
Fiber optic RS232 cable, 9-pin D-sub |
|
Remote operation |
All key functions are remotely operated by ESM-100 graphics software through fiber optic connection to a computer |
|
Temperature and humidity operating range |
-20~50 °C; 0-90% relative humidity, non condensing |
|
size |
365mm x 83mm x 56mm |
|
weight |
570g |