Druck multifunctional process signal calibrator from the United States

(DPI880)

Product features:

1.The Druck DPI880 multifunctional process signal calibrator from the United States is a compact, easy-to-use tool that can be used for testing, configuring, and calibrating almost all process parameters. It features a large backlit LCD display, menu driven interface, sturdy and weather resistant design, compact structure, simple operation, and easy portability

2. Measure and output mA, mV,V, Resistance, frequency, and pulse signal

3.Measurement and simulation output of 8 types of RTDs and 12 types of thermocouples

4.32Optional pressure module with a range from 2.5Kpa to 70Mpa

5.The function of simultaneously reading dual readings

6.Automatic switch test and pressure leakage test

7.1000Data storage function with real-time time tags

8.Can provide 24V circuit power supply

9.Support IDOS pressure module, plug and play

10.Automatic switch detection: Capture the opening/closing of the gate for quick and effective safety system checks

11.HartResistance: It can be switched to the circuit when needed by the Hart digital communicator, thus avoiding the inconvenience of carrying a 250 Ω resistor

12.Automatic detection of 2-wire, 3-wire, and 4-wire: Quickly detect faulty sensors and wiring

function

Measurement or simulation

IDOS

mA

V

mV

RTD

TC

V

mA

correct

correct

mA(24V)

V

Switch test

Support IDSO pressure module

leak test

correct

Electrical signal indicators:

measure

accuracy

0 ~ 120.00mV

0.02%rdg+2counts

0 ~ 30.000V

0.03%rdg+2counts

0 ~ 55.000mA

0.02%rdg+3counts

0 ~ 4000.0Ω

0.1 ~ 1.3Ω

output

accuracy

0 ~ 120.00mV

0.02%rdg+2counts

0 ~ 12.000V

0.02%rdg+2counts

0 ~ 24.000mA

0.02%rdg+2counts

0 ~ 4000.0Ω

0.1 ~ 1.3Ω

frequency index

measure

accuracy

0 ~ 999.999Hz

0.003%rdg+2counts

0 ~ 50.0000KHz

0.003%rdg+2counts

0 ~ 999999cpm/cph

0.003%rdg+3counts

output

accuracy

0 ~ 999.99Hz

0.003%rdg+0.0023Hz

0 ~ 50.000KHz

0.003%rdg+0.0336KHz

0 ~ 59999cpm

0.003%rdg+0.138cpm

0 ~ 99999cph

0.003%rdg+0.5cph

Voltage input

30V Max

trigger voltage

0 ~ 12VResolution 0.1V

AMPLITUDE

0 ~ 12V DC

0 ~ 12V AC

Thermal resistance index

Measurement/simulation output

standard

range

accuracy

Pt50(385)

IEC 751

-200 ~ 850

0.5

Pt100(385)

IEC 751

-200 ~ 850

0.25

Pt200(385)

IEC 751

-200 ~ 850

0.6

Pt500(385)

IEC 751

-200 ~ 850

0.4

Pt1000(385)

IEC 751

-200 ~ 400

0.2

D100(392)

JIS 1604-1989

-200 ~ 650

0.25

Ni100

DIN 43760

-60 ~ 250

0.2

Ni120

MINCO 7-120

-80 ~ 260

0.2

Ohms

0 ~ 4000Ω

0.1 ~ 1.3Ω

 

Thermocouple indicators

Measurement/simulation output

standard

range

accuracy

K

IEC 584

-270 ~ 1370

0.6

J

IEC 584

-210 ~ 1200

0.5

T

IEC 584

-270 ~ 400

0.3

B

IEC 584

50 ~ 1820

1.0

R

IEC 584

-50 ~ 1769

1.0

S

IEC 584

-50 ~ 1769

1.4

E

IEC 584

-270 ~ 1000

0.4

N

IEC 584

-270 ~ 1300

0.6

L

DIN 43710

-200 ~ 900

0.3

U

DIN 43710

-200 ~ 600

0.3

C

0 ~ 2320

1.0

 

D

0 ~ 24950

1.0

 

mV

-10mV ~ 100mV

0.2%rdg

 

Item number

product description

DPI 880

Multi functional process signal calibrator

Druck multifunctional process signal calibrator from the United States

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!