Druck multifunctional process signal calibrator from the United States
(DPI880)
Product features:
1.The Druck DPI880 multifunctional process signal calibrator from the United States is a compact, easy-to-use tool that can be used for testing, configuring, and calibrating almost all process parameters. It features a large backlit LCD display, menu driven interface, sturdy and weather resistant design, compact structure, simple operation, and easy portability
2. Measure and output mA, mV,V, Resistance, frequency, and pulse signal
3.Measurement and simulation output of 8 types of RTDs and 12 types of thermocouples
4.32Optional pressure module with a range from 2.5Kpa to 70Mpa
5.The function of simultaneously reading dual readings
6.Automatic switch test and pressure leakage test
7.1000Data storage function with real-time time tags
8.Can provide 24V circuit power supply
9.Support IDOS pressure module, plug and play
10.Automatic switch detection: Capture the opening/closing of the gate for quick and effective safety system checks
11.HartResistance: It can be switched to the circuit when needed by the Hart digital communicator, thus avoiding the inconvenience of carrying a 250 Ω resistor
12.Automatic detection of 2-wire, 3-wire, and 4-wire: Quickly detect faulty sensors and wiring
function |
Measurement or simulation |
IDOS |
||||||
mA |
V |
mV |
RTD |
TC |
V |
|||
mA |
correct |
correct |
||||||
mA(24V) | ||||||||
V | ||||||||
Switch test | ||||||||
Support IDSO pressure module |
||||||||
leak test |
correct |
|||||||
Electrical signal indicators:
measure |
accuracy |
0 ~ 120.00mV |
0.02%rdg+2counts |
0 ~ 30.000V |
0.03%rdg+2counts |
0 ~ 55.000mA |
0.02%rdg+3counts |
0 ~ 4000.0Ω |
0.1 ~ 1.3Ω |
output |
accuracy |
0 ~ 120.00mV |
0.02%rdg+2counts |
0 ~ 12.000V |
0.02%rdg+2counts |
0 ~ 24.000mA |
0.02%rdg+2counts |
0 ~ 4000.0Ω |
0.1 ~ 1.3Ω |
frequency index
measure |
accuracy |
0 ~ 999.999Hz |
0.003%rdg+2counts |
0 ~ 50.0000KHz |
0.003%rdg+2counts |
0 ~ 999999cpm/cph |
0.003%rdg+3counts |
output |
accuracy |
0 ~ 999.99Hz |
0.003%rdg+0.0023Hz |
0 ~ 50.000KHz |
0.003%rdg+0.0336KHz |
0 ~ 59999cpm |
0.003%rdg+0.138cpm |
0 ~ 99999cph |
0.003%rdg+0.5cph |
Voltage input |
30V Max |
trigger voltage |
0 ~ 12VResolution 0.1V |
AMPLITUDE |
0 ~ 12V DC |
0 ~ 12V AC |
Thermal resistance index
Measurement/simulation output |
standard |
range |
accuracy |
Pt50(385) |
IEC 751 |
-200 ~ 850℃ |
0.5℃ |
Pt100(385) |
IEC 751 |
-200 ~ 850℃ |
0.25℃ |
Pt200(385) |
IEC 751 |
-200 ~ 850℃ |
0.6℃ |
Pt500(385) |
IEC 751 |
-200 ~ 850℃ |
0.4℃ |
Pt1000(385) |
IEC 751 |
-200 ~ 400℃ |
0.2℃ |
D100(392) |
JIS 1604-1989 |
-200 ~ 650℃ |
0.25℃ |
Ni100 |
DIN 43760 |
-60 ~ 250℃ |
0.2℃ |
Ni120 |
MINCO 7-120 |
-80 ~ 260℃ |
0.2℃ |
Ohms |
0 ~ 4000Ω |
0.1 ~ 1.3Ω |
|
Thermocouple indicators
Measurement/simulation output |
standard |
range |
accuracy |
K |
IEC 584 |
-270 ~ 1370℃ |
0.6℃ |
J |
IEC 584 |
-210 ~ 1200℃ |
0.5℃ |
T |
IEC 584 |
-270 ~ 400℃ |
0.3℃ |
B |
IEC 584 |
50 ~ 1820℃ |
1.0℃ |
R |
IEC 584 |
-50 ~ 1769℃ |
1.0℃ |
S |
IEC 584 |
-50 ~ 1769℃ |
1.4℃ |
E |
IEC 584 |
-270 ~ 1000℃ |
0.4℃ |
N |
IEC 584 |
-270 ~ 1300℃ |
0.6℃ |
L |
DIN 43710 |
-200 ~ 900℃ |
0.3℃ |
U |
DIN 43710 |
-200 ~ 600℃ |
0.3℃ |
C |
0 ~ 2320℃ |
1.0℃ |
|
D |
0 ~ 24950℃ |
1.0℃ |
|
mV |
-10mV ~ 100mV |
0.2%rdg |
|
Item number |
product description |
DPI 880 |
Multi functional process signal calibrator |
Druck multifunctional process signal calibrator from the United States