Druck DPI602S pressure indicator
Pressure indicator/pressure calibrator - DPI 800/DPI 802 | Pressure calibrator DPI 800/DPI 802
The DPI800 series pressure calibrator is a newly launched handheld calibrator by General Electric Sensing and Measurement, which is full range, high reliability, and easy to use, providing users with the best solution for process parameter testing and calibration. The enhanced performance and technological innovation of the DPI800 series pressure calibrator are sure to bring higher efficiency to your work and earn your trust. Performance and advantages of DPI800 series pressure calibrator
The DPI800 series pressure calibrator is an ideal tool for pressure measurement and calibration, with a pressure range from 2.5kPa to 70Mpa, including vacuum range; No matter how the environmental conditions change, the accuracy meets the indicators within one year.
- Range: 2.5kPa~70MPa
- Internal single/dual range optional
- Stress testing and maintenance
- Calibration of pressure transmitter
- Circuit setting and diagnosis
- Switch quantity test
DPI800 series pressure calibrator service types
- Calibration: Offline, on-site
Repair: Offline, on-site
- evaluate
|
Selection of pressure range |
G/D |
G |
A |
Media+- |
overvoltage |
Accuracy (% FS) |
|
Standard height |
||||||
|
±7/20/35/70KPa |
● |
|
|
②③ |
2×FS |
0.0750.03 |
|
0 - 35KPa |
|
|
● |
② |
2×FS |
0.1N/A |
|
Negative 100- (100/200KPa) |
● |
|
|
②③ |
2×FS |
0.050.01 |
|
0 - 200KPa |
|
|
● |
② |
2×FS |
0.075N/A |
|
Negative 100- (350/500/700KPa) |
|
● |
|
① |
2×FS |
0.050.1 |
|
Negative 0.1- (1/2MPa) |
|
● |
|
① |
2×FS |
0.050.1 |
|
0 - (0.7/2MPa) |
|
|
● |
① |
2×FS |
0.75N/A |
|
0 - (3.5/7/10/13.5/20MPa) |
|
● |
|
① |
2×FS |
0.050.01 |
Thermal resistance indicator/calibrator - DPI 811/812
DPI 811/812 is an ideal tool for measuring and simulating RTD sensors, as well as for inspecting RTD probes, indicators, recorders, and controllers. Supports 8 standard RTDs.
·Measure/simulate output thermal resistance signal
·2, 3, 4 wire automatic detection system
·Temperature testing and maintenance
·Transmitter calibration
·Circuit setting and diagnosis
·Switch quantity test
·Support IDOS pressure module
|
Measurement/output simulation |
standard |
range |
precision |
|
PT50(385) |
IEC 751 |
Negative 200-850 ℃ |
Negative 0.5 ℃ |
|
PT100(385) |
IEC 751 |
Negative 200-850 ℃ |
0.25℃ |
|
PT200(385) |
IEC 751 |
Negative 200-850 ℃ |
0.6℃ |
|
PT500(385) |
IEC 751 |
Negative 200-850 ℃ |
0.4℃ |
|
PT1000(385) |
IEC 751 |
Negative 200-400 ℃ |
0.2℃ |
|
D100(392) |
JIS1604-1989 |
Negative 200-650 ℃ |
0.25℃ |
|
Ni100 |
DIN 43760 |
Negative 60-250 ℃ |
0.2℃ |
|
Ni120 |
MINCO 7-120 |
Negative 80-260 ℃ |
0.2℃ |
|
0hms |
|
0-4000Ω |
0.1 - 1.3Ω |
Thermocouple indicator/calibrator DPI 821/822
DPI 821/822 measurement and simulation sensors are ideal tools for checking temperature probes, indicators, recorders, and controllers.
·Measure/simulate thermocouple signal output
·Unique CJ cold end compensation technology
·Temperature testing and maintenance
·Transmitter calibration
·Circuit setting and diagnosis
·Switch quantity test
·Support IDOS pressure module
|
type |
standard |
range |
precision |
|
K |
IEC 584 |
Negative 270-1370 ℃ |
0.6℃ |
|
J |
IEC 584 |
Negative 210-1200 ℃ |
0.5℃ |
|
T |
IEC 584 |
Negative 270-400 ℃ |
0.3℃ |
|
B |
IEC 584 |
50 - 1820℃ |
1.0℃ |
|
R |
IEC 584 |
Negative 50-1769 ℃ |
1.0℃ |
|
S |
IEC 584 |
Negative 50-1769 ℃ |
1.4℃ |
|
E |
IEC 584 |
Negative 270-1000 ℃ |
0.4℃ |
|
N |
IEC 584 |
Negative 270-1300 ℃ |
0.6℃ |
|
L |
DIN 43710 |
Negative 200-900 ℃ |
0.3℃ |
|
U |
DIN 43710 |
Negative 200-600 ℃ |
0.3℃ |
|
C |
|
0 - 2320℃ |
1.0℃ |
|
D |
|
0 - 2495℃ |
1.0℃ |
|
mV |
|
Negative 10mV -100mV |
|
Dual channel loop calibrator DPI 832
DPI 832 can measure and output mA, mV, V signals, and can capture the values of switch actions. It is an ideal tool for circuit debugging, transmitter maintenance debugging, DCS, PLC input and output card debugging.
·Dual channel reading capability: outputting mA, mV, V signals, and can simultaneously measure mA or capture switch action values
·Measure and output mA, mV, V signals
·Can provide 24V circuit power supply
·Large backlit LCD display, menu driven interface
·Durable and weatherproof design
·Compact structure, simple operation, easy to carry
·Support IDOS pressure module
|
measurement |
precision |
|
0 ~ 120.00mV |
0.02%rdg+2counts |
|
0 ~ 30.00V |
0.03%rdg+2counts |
|
0 ~ 55.00000mA |
0.02%rdg+3counts |
|
output |
|
|
0 ~ 120.00mV |
0.02%rdg+2counts |
|
0 ~ 12.00V |
0.02%rdg+2counts |
|
0 ~ 24.00mA |
0.02%rdg+2counts |
Frequency calibrator DPI 841/842
For process technology and electronics engineers, DPI 841/842 is an ideal tool that provides high-precision calibration standards and versatile testing tools capable of measuring Hz, kHz, CPM, CPH, and pulses. Its * functionality makes debugging and maintenance of electrical signal and frequency instruments simple, including tachometers, frequency meters, batch counters, flow meters, and more.
·High performance frequency measurement and generator capable of generating three different wave modes with amplitudes ranging from 0.1 to 24V
·Measurement and occurrence frequency from 0.01 Hz to 50 kHz
·CPM, CPH, and total counter
·Sine, square, and triangular wave patterns
·MA measurement, switch testing, and can provide 24V circuit power supply
·Large backlit LCD display, menu driven interface
·Durable and weatherproof design
·Compact structure, simple operation, easy to carry
·Support IDOS pressure module
|
measurement |
precision |
|
0 ~ 999.999Hz |
0.003%rdg+2counts |
|
0 ~ 50.0000KHz |
0.003%rdg+2counts |
|
0 ~ 999999cpm/cph |
0.003%rdg+2counts |
|
output |
|
|
0 ~ 999.99Hz |
0.003%rdg+0.0023Hz |
|
0 ~ 50.000kHz |
0.003%rdg+0.0336KHz |
|
0 ~ 99999cpm |
0.003%rdg+0.138cpm |
|
0 ~ 99999cph |
0.003%rdg+0.5cph |
Multi functional process signal calibrator DPI 880
DPI 880 is a compact, easy-to-use tool that can be used for testing, configuring, and verifying almost all process parameters. It can measure and output mA, mV, V, RTD, thermocouple, resistance, frequency, pulse, and pressure, and can grasp the value of switch action. It can provide 24V power supply for the circuit.
·Measure and output mA, mV, V, resistance, frequency, and pulse signals
·Measurement and simulation output of 8 types of RTDs and 12 types of thermocouples
·The function of simultaneously reading dual readings
·Automatic switch test and pressure leakage test
·1000 point data storage function with real-time time tags
·Can provide 24V circuit power supply
·Large backlit LCD display, menu driven interface
·Durable and weatherproof design
·Compact structure, simple operation, easy to carry
·Support IDOS pressure module
DPI 880 electrical signal indication
|
measurement |
precision |
|
0 ~ 120.00mV |
0.02%rdg+2counts |
|
0 ~ 30.000V |
0.03%rdg+2counts |
|
0 ~ 55.000mA |
0.02%rdg+3counts |
|
0 ~ 4000.0Ω |
0.1 ~ 1.3Ω |
|
output |
|
|
0 ~ 120.00mV |
0.02%rdg+2counts |
|
0 ~ 12.000V |
0.02%rdg+2counts |
|
0 ~ 24.000mA |
0.02%rdg+2counts |
|
0 ~ 4000.0Ω |
0.1 ~ 1.3Ω |
DPI 880 frequency index
|
measurement |
precision |
|
0 ~ 999.999Hz |
0.003%rdg+2counts |
|
0 ~ 50.0000KHz |
0.003%rdg+2counts |
|
0 ~ 999999cpm/cph |
0.003%rdg+3counts |
|
output |
|
|
0 ~ 999.99Hz |
0.003%rdg+0.0023Hz |
|
0 ~ 50.000KHz |
0.003%rdg+0.0336KHz |
|
0 ~ 59999cpm |
0.003%rdg+0.138cpm |
|
0 ~ 99999cph |
0.003%rdg+0.5cph |
|
Voltage input |
30V Max |
|
trigger voltage |
0 ~ 12V, Resolution 0.1V |
|
AMPLITUDE |
0 ~ 12V DC |
|
0 ~ 12V AC |
DPI 880 thermal resistance index
|
Measurement/simulation output |
standard |
range |
precision |
|
Pt50 (385) |
IEC 751 |
-200 ~ 850℃ |
0.5℃ |
|
Pt100 (385) |
IEC 751 |
-200 ~ 850℃ |
0.25℃ |
|
Pt200 (385) |
IEC 751 |
-200 ~ 850℃ |
0.6℃ |
|
Pt500 (385) |
IEC 751 |
-200 ~ 850℃ |
0.4℃ |
|
Pt1000 (385) |
IEC 751 |
-200 ~ 400℃ |
0.2℃ |
|
D100 (392) |
JIS 1604-1989 |
-200 ~ 650℃ |
0.25℃ |
|
Ni100 |
DIN 43760 |
-60 ~ 250℃ |
0.2℃ |
|
Ni120 |
MINCO 7-120 |
-80 ~ 260℃ |
0.2℃ |
|
Ohms |
|
0 ~ 4000Ω |
0.1 ~ 1.3Ω |
DPI 880 thermocouple specifications
|
Measurement/simulation output |
standard |
range |
precision |
|
K |
IEC 584 |
-270 ~ 1370℃ |
0.6℃ |
|
J |
IEC 584 |
-210 ~ 1200℃ |
0.5℃ |
|
T |
IEC 584 |
-270 ~ 400℃ |
0.3℃ |
|
B |
IEC 584 |
50 ~ 1820℃ |
1.0℃ |
|
R |
IEC 584 |
-50 ~ 1769℃ |
1.0℃ |
|
S |
IEC 584 |
-50 ~ 1769℃ |
1.4℃ |
|
E |
IEC 584 |
-270 ~ 1000℃ |
0.4℃ |
|
N |
IEC 584 |
-270 ~ 1300℃ |
0.6℃ |
|
L |
DIN 43710 |
-200 ~ 900℃ |
0.3℃ |
|
U |
DIN 43710 |
-200 ~ 600℃ |
0.3℃ |
|
C |
|
0 ~ 2320℃ |
1.0℃ |
|
D |
|
0 ~ 24950℃ |
1.0℃ |
|
mV |
|
-10mV ~ 100mV |
0.2%rdg |