product overview
The DL2030 cyanide automatic online analyzer uses a combination of online distillation absorption technology and sequential injection spectrophotometry to detect the concentration of cyanide in water. This instrument can achieve real-time, rapid, and accurate analysis of cyanide concentration in water.
application area
Surface water, tap water, municipal sewage, industrial wastewater and other water bodies.
Technical Parameter
project |
TECHNICAL INDEX |
Analysis method |
isonicotinic acid-Barbituric acid spectrophotometric method |
measuring range |
0~2mg/L (Can be adjusted as needed) |
Indication error |
±5% |
precision |
5% |
zero drift |
±5% |
Span Drift |
±5% |
detection limit |
0.001 mg/L |