Main features
Infrared calibration transmission source, standard transmission radiation pyrometer
• Transfer source DS10Ncal
Calibration temperature range: 600~2500 ° C or 3000 ° C
• Wavelength 0.8~1.1 μ m
Performance
Digital monochrome infrared thermometer, temperature measurement range: 200~3000 ° C
• Short response time, minimum 2 ms
• Minimum measurable diameter of 1.2mm
Communication interface RS485
On site display and on-site button adjustment parameters
Variable focal length with aiming light or lens for aiming
TECHNICAL INDEX
model |
DS10Ncal |
Measuring range |
600 ~2500 °C |
spectral range |
0.8μm~1.1μm |
LENS |
Fixed focal length, minimum measurement target 3mm@600mm |
Distance coefficient ratio |
200:1 |
temperature measurement error |
Refer to the factory inspection form |
repeatability |
0.1K(Tamb= 23 °C, ε = 1, t95 = 1s) |
NETD |
0.1K(TU= 23 °C, ε = 1, t95 = 1s) |
Response time (t)95) |
Fixed at 1 second |
emissivity |
0.900~1.100, adjustable through communication interface or infrared thermometer button |
output signal |
0/4 ~ 20 mA, Adjustable through software, temperature linear, maximum load: 500 Ω |
communication interface |
Connect via the included USB cable |
sighting system |
Aim through a lens |
Adjustable parameters |
Adjustable through software: SSE correction (emissivity), temperature unit in ° C or ° F |
power supply |
Wall mounted 100~240VAC |
power consumption |
Maximum 1.5 W |
operate temperature |
23 °C ± 3 K |
storage temperature |
0~70 °C |
weight |
Approximately 520 grams |
size |
54 × 54 mm, length 170 mm |
protection grade |
IP65 compliant with DIN EN 60529 and DIN 40050 |
test specification |
EN 55 011:1998, class A |
CE certification |
Compliant with EU regulations (EN 50 011) |
scope of supply |
PYROSPOT DS10Ncal, Installation base with tripod adapter, operation manual, 10 point inspection sheet (traceable to PTB standard), software PYROSOFT Spot (no connection cable, please order separately), radiation source size effect (SSE) curve, various globally compatible plug adapters (EU/US/UK/AU) USB connection cables, plastic packaging box |
Ultra high precision
DS10Ncal_DG10Ncal | ||
Radiation temperature ° C |
Actual measurement error ° C |
% |
299.95 |
0.21 |
0.07% |
450.009 |
0.078 |
0.02% |
600.026 |
0.079 |
0.01% |
750.02 |
0.12 |
0.02% |
899.97 |
0.15 |
0.02% |
1050.1 |
0.83 |
0.08% |
1197.27 |
0.89 |
0.07% |
1397.5 |
1 |
0.07% |
1796.9 |
1.2 |
0.07% |
List of DIAS Black body Furnace Models:
Infrared standard transfer source Transfer Pyrometer for DIAS blackbody furnace