Performance characteristics
Wide adaptability range, analog sensor technology, and easier matching measurement.
Stable 4-20mA transmission mode, wider adaptability, and easier connection!
● Signal output: 4-20mA current.
●Application scope: Various water treatments (purified water/drinking water/surface water/various water supplies/industrial wastewater treatment), aquaculture, environmental monitoring, CIP and other industries.
●Signal output: 4-20mA (2-wire analog signal), easy to connect to third-party devices such as PLC, DCS, industrial control computer, universal controller, paperless recording instrument or touch screen.
●Easy to install: 3/4 inch NPT thread (pipe thread), easy to install in pipelines and tanks. The probe and display part can be separated and connected by cables.
Technical Parameter
|
range resolution |
0.01~20μS/cm 0.01 0.1~200μS/cm 0.1 0~5000μS/cm 1 0~200mS/cm 0.1 |
accuracy |
±1.5%F.S. |
operation temperature |
0~65℃ |
Working pressure |
<0.6MPa |
power supply |
24VDC±10% |
signal output |
4-20mA two-wire system (isolated signal) |
Liquid receiving material |
ABS/SUS316L (default)/PVDF/PVC (customizable) |
Installation method |
3/4 "NPT thread |
Maintenance |
It is recommended to clean the small amount once a month. |
Display mode |
4-digit LED display screen |
Measurement compensation |
Automatic temperature compensation (PT1000) |
size |
45 × 50 × 190mm (there may be slight differences in different subdivision models) |
weight |
170g (there may be slight differences among different sub models) |
Calibration method |
Two-point calibration |
protection grade |
IP65 |
measuring principle |
Constant 1.0, 0.1, 0.01 Conductivity electrode: (contact type) Two pole electrode measurement principle Electromagnetic conductive electrode: (non-contact) electromagnetic waves passing through a conductive liquid medium |
brand |
DOWESTON |
DOCUMENTATION
- Propagandist Manual
- Product manual
- Drawings and software