Honeywell Online in the United StatesConductivity analyzerUDA2182+Conductivity electrode04973
Conductivity analyzerUDA2182-CC1-NN2-NN-N-0000
Conductivity meter UDA2182-CC1-CC2-NN-N-0000
Conductivity electrode 04973-X10-333-20-00-000
Conductivity electrode 04973-XX1-333-20-00-000
Conductivity electrode 04973-X01-333-20-00-000
Conductivity electrode 04973-001-333-20-00-000
Conductivity electrode5000TC-00-1-000
Conductivity electrode5000TC-00-2-000
Conductivity electrode5000TC-00-3-000
Conductivity electrode5000TC-00-4-000
Honeywell online acid-base concentration analyzer from the United StatesUDA2182+acidity concentration electrode 04905
Acid concentration electrode 04905-X50-44-333-20-000-000
Alkali concentration electrode 04905-X50-33-333-20-000-0000
Acid concentration junction box04905-X50-44-333-X1-000-000
alkaliConcentration junction box04905-X50-33-333-X1-000-0000
Acid concentration electrodeTF-20-K50-H
Alkaline concentration electrodeTF-20-K50-N
The Honeywell UDA2182 pH meter calculates pH by measuring the specific conductivity and positive conductivity of ammonia based pure water, and its working principle complies with the international VBG450 standard. In a pure water environment, pH measurement is affected by factors such as unstable liquid contact potential, static charge, grounding conditions, temperature changes, glass electrode quality, and external interference. The corresponding measures taken can only reduce the impact, but cannot completely eliminate it. Therefore, pure water pH measurement has always been a global problem, manifested in large deviations in pH meter readings, instability, and sensitivity to flow rate.
The Honeywell UDA2182PD pH meter uses conductivity measurement method, effectively avoiding the various drawbacks of pH electrode measurement.
The entire system includes instruments, electrodes, flow tanks, cation exchange columns, valves, flow meters, and corresponding pipelines. All components are concentrated on a stainless steel plate.
The instrument has two electrodes, one with an electrode constant of 0.1 cm-1 for specific conductivity measurement, and the other with an electrode constant of 0.01 cm-1 for positive conductivity measurement.
As shown in the figure on the right, the sample water enters the flow tank 1 through the inlet valve to measure the specific conductivity, then enters the cation exchange column to remove all alkalizing reagents, enters the second conductivity electrode to measure the conductivity, and finally the sample water is discharged after passing through the flowmeter.
Application measurement points:
Suitable for pH measurement of pure water containing one or several alkalizing agents and ammonia added pure water, with a positive conductivity<0.3 μ S/cm, pH 7.5-10.5, and measuring points including feedwater, condensate, steam, internal cooling water, etc.
Working Principle:
The instrument simultaneously measures the specific conductivity value (sample water conductivity) and the cation conductivity value (conductivity exchanged with cation resin) of the sample water, and calculates the pH value of the sample water by the difference between the two.
Features:
1. Less interference factors, high accuracy, AVT water quality error<0.01pH, and stable measurement values
2. Low maintenance requirement, usually does not require calibration, only needs to clean the electrode once a year
3. No consumables, low operating costs
4. Can simultaneously measure pH, specific conductivity, and positive conductivity
5. Suitable for adding one or more types of water such as ammonia, hydrazine, morpholine, ethanolamine, etc
6. Multiple nonlinear temperature compensations are available for different water qualities
7. All components such as the cation exchange column are concentrated on a stainless steel plate, which is beautiful and easy to install
8. Long life titanium electrode, electrode constant accurate to one ten thousandth, high measurement accuracy
9. No pH electrode, no need for pH measurement calibration, instrument maintenance free, no consumption
10. Overall wall mounted, easy to install and maintain
Technical Specifications:
display |
Color LED with background light Screen size: 66.8mm (W) x35.5mm (H) Lattice number: 128 (W) x64 (H) Can display two conductivity values, pH, and sample water temperature simultaneously |
measurement range |
Conductivity: 0.055-200 μ S/cm pH:7.5~10.5pH |
keyboard |
10 single keys, wear-resistant |
Shell material |
GE Valox357 |
precision |
0.5% of the reading |
output accuracy |
±0.01mA |
repetitiveness |
0.05% |
temperature accuracy |
± 0.1 ℃ at -10~100 ℃, ± 1 ℃ at 101~140 ℃ |
Operating temperature |
0~60℃ |
relative humidity |
5-90%, no condensation within 40 ℃ (104 ℉) |
analog output |
2-channel 0/4~20mA (standard), optional third channel, maximum load 750 Ω |
Communication output |
Ethernet/RS485 Modbus/Profibus optional |
temperature compensation |
NaCl (pure water) NH3 (ammonia added pure water), HCl (positive conductivity), PO-4 (phosphate containing boiler water), C4H9NO (morpholine added pure water), etc |
power supply |
90-264VAC, 47-63Hz, 15VA, E ² PROM power-off memory |
Safety Certification |
UL/CSA/CE FM, Class 1, Div2 |
Overall dimensions |
850(H)x 280(W)x 202.5(D)mm |
installation dimensions |
824 (H) x 254 (W) mm, 4- Ø 10mm mounting hole |
Sample water inlet size |
1/4 "card sleeve connector |
Size of sample water discharge outlet |
3/4 "NPT external thread |
electrical interface |
4x1/2 "wiring holes |
Protection level |
CSA Type 4X(NEMA 4X)/IP66 外壳 |
install |
Wall mounted |