The technical specifications of this series of instruments meet the requirements of the JJF1190-2008 calibration regulations issued by the State Administration of Metrology. The entire machine adopts American microcomputer control processing technology and the most advanced SMT chip surface mount packaging technology, semiconductor laser sensor technology, and noise proof air pump internationally, and can directly print test results. It has the characteristics of multiple functions, high measurement accuracy, fast speed, easy portability, and simple operation. The instrument can simultaneously measure the number of dust particles of multiple sizes in one sampling, and can choose to observe the number and changes of particles of a certain size, which is very convenient for researching, detecting, and evaluating various clean environments. This series of instruments features advanced performance design, stable and reliable quality
This series of products has been widely used in clean room testing; On site inspection and leak detection of filters; It can monitor ultra clean workbenches, biosafety cabinets, HVAC systems, computer rooms, beverage packaging environments, medical device production environments, hospital clean operating rooms, automotive spraying environments, microelectronics, pharmaceuticals, biochemical products, food hygiene, fine chemicals, precision machinery, aerospace and other production and research departments. It is the preferred instrument for pharmaceutical companies and their supervisory management departments to implement GMP standards and for electronic production enterprises.

Main technical parameters and performance:
1. Light source: Full semiconductor laser light source
2. Sampling volume: 2.83L/min (0.1cfm/min)
3. Detection range: 100 to 1 million levels
4. Allow the tested air to have a dust concentration of no more than 100000 particles/2.83L
5. Particle size channel: 0.3 0.5 1.0 3.0 5.0 10.0 (μ m) six levels
6. Display or print can convert particles contained within 2.83 liters/minute into particles contained within 1m3.
7. Sampling period: 1-10 (min)
8. Self purification time: ≤ 15 (min)
9. Calibration: Traceable to the National Institute of Standards and Technology (NIST) in the United States. Our company has passed the national metrology standard building assessment and can be traced back to the Shanghai Institute of Metrology and Measurement Technology for self calibration or third-party national metrology institutions for calibration
10. Working environment: Temperature: 10-35 ℃ Relative temperature: 20-75% RH
11. Power consumption: 25W
12. Range and accuracy of measuring temperature and humidity: (optional)
(1) Temperature: 0-50 ℃± 1 ℃
(2) Humidity: 0-100% RH ± 5%
13. Sampling points set from 2 to 7 points
14. Set the sampling frequency to 2-9 times per point
UCL report: compliant with FS-209E and Chinese GMP standards
16. Working hours: 8 hours
17. Power supply: AC220V ± 10% 50 ± 2Hz
18. Weight: 4.8kg
19. Dimensions: 260 × 130 × 340
20. Simultaneously detect the concentration of dust particles in six different sizes, and display them in numerical or self selected sizes.
21. Remarks: Built in printer, automatic determination of purification level, power sampling head, sampling rack
(Temperature and humidity probe is optional)