1、 Product Introduction: The main function of CH-0232 Liquefied Gas Copper Corrosion Tester is to evaluate the corrosion of copper metal by liquefied
Product details

1、 Product Introduction

CH-0232 Liquefied Gas Copper Corrosion TesterThe main function is to evaluate the corrosion detection of copper metal by liquefied gas or other gaseous fuels. This type of testing is important for ensuring the quality of fuel and preventing

Equipment damage caused by corrosion is crucial. The working principle of the liquefied gas copper corrosion tester usually involves placing a standard sized copper sheet in the liquefied gas sample to be tested, under specified conditions

After maintaining at a certain temperature for a period of time, the degree of corrosion and the extent of corrosion can be determined by comparing the appearance changes, such as color changes, of the copper sheet before and after corrosion using a color chart. This test helps

Ensure the safety and reliability of liquefied gas.

Operation steps of copper corrosion tester

1. On the premise of ensuring the presence of water, turn on the measuring instrument and wait for it to stabilize at 40 ℃;

2. Open the bottom valve B of the test cylinder and inject about 1mL of distilled water into the clean test cylinder, then rotate it to wet the cylinder wall and allow the residual liquid to be discharged from valve B.

3. Use tweezers to clamp the newly polished copper sheet, immediately hang it onto the hook of the cylinder, and place it into the cylinder (keeping the cylinder vertical). Ensure that the bottom edge of the copper sheet is at least 6mm away from the bottom of the cylinder during assembly.

After the instrument is assembled, close valves A and B.

4. Keep the test cylinder in a vertical position to prevent the copper sheet from getting wet with water, and tightly connect the sample source and the A valve of the test cylinder with the connecting hose and its accessories that have been washed clean by the sample.

First, open the valve on the sample source, and then open valve A on the cylinder to allow some samples to enter the cylinder.

5. Close valve A and do not detach the test cylinder from the sample source. Reverse the test cylinder and open valve B to remove the air from the test cylinder, then close valve B. Then turn the test cylinder back to the vertical position,

Open valve B to drain all residual liquid. Immediately close valve B in the vertical position and open valve A to fill the test cylinder with the sample. When the test cylinder is full, close the valve of the sample source and valve A,

Remove the connecting hose.

6. When the connecting hose is just removed and the cylinder is in an upright position, immediately slightly open valve A to remove the liquid above the end of the immersion tube from the test cylinder. At the beginning of the gas

When valve A appears, close valve A.

7. Immediately immerse the cylinder in a constant temperature water bath at 40 ℃ and let it sit in the bath for 1 hour ± 5 minutes.

8. At the end of the experiment, remove the cylinder from the water bath, place it in an upright position, open valve B, and discharge the liquid and most of the gas.

9. When there is only a slight pressure in the cylinder, immediately disconnect the device and compare the copper sheet treated with liquefied petroleum gas with the corrosion standard color plate. When comparing, the test piece should be compared with the standard

Place the color palette in the direction of light reflection at an angle of approximately 45 ° for observation.


2、 Technical parameters

1. Working voltage: AC220V ± 10 50Hz

2. Power: 2000W3. Temperature control range: room temperature~100 ℃

4. Temperature sensor: Pt100

5. Temperature control method: digital display temperature controller, temperature control accuracy: 0.1 ℃;

6. Stainless steel bath and shell, rust proof

Compliant with the Guo standard:

SH/T 0232

D130/IP154

GB/T5096




3、 After sales service commitment 

1. Warranty period: The instrument is covered by a 12-month warranty from the date of self installation, commissioning, and acceptance, and requires personal maintenance.

2. In case of instrument malfunction, timely response or telephone guidance for maintenance can be provided, and professional technical service engineers can be dispatched to the site within 48 hours to troubleshoot.

3. If there is a software upgrade for the same series of instruments, our company will provide free upgrade services for users.

4. Installation and Debugging: After the user has prepared all the installation conditions and requested installation, our company will send professional engineers to install and debug on-site for free, providing free services to the user

Conduct centralized on-site training for users on the use, maintenance, and upkeep of the instrument until they are proficient in operating it independently.