VIP member
Battery pack simulation high-altitude test box GX-3020-ZC12M
Product Name: GX-3020-ZC12M Battery Pack Simulated High Altitude Test Chamber Product Model: GX-3020-ZC12M Product Usage: The battery pack simulated h
Product details
1、 Product Name:Battery pack simulation high-altitude test box
2、 Product model: GX-3020-ZC12M
3、 Equipment Description:Battery pack simulation high-altitude test boxMainly used to simulate and evaluate the safety performance of PACK level battery packs and systems under vacuum low voltage (i.e. high altitude) conditions.High altitude battery pack testing machineSimulate that the battery is put into the low air pressure box, adjust the air pressure in the test box to the air pressure condition with an altitude of 4000m or the same height, the equipment can automatically control the test cycle (such as 5H), monitor the air pressure change in the box throughout the whole process, automatically keep the air pressure in the box constant, and realize the automatic termination of the test.
Present the actual usage status of battery packs or systems in the laboratory, identify and solve problems, optimize product design, and improve product quality.
| GB/T 31467.3-2015 |
7.12 High altitude testing for lithium-ion power battery packs and systems for electric vehicles - Part 3: Safety requirements and testing methods
|
| GB/T 31485-2015 | 6.2.11&6.3.11 Low pressure testing of power batteries for electric vehicles - Safety requirements and test methods |
| GB/T 31486-2015 | Requirements and Test Methods for Electrical Performance of Power Batteries for Electric Vehicles |
| GB 31241-2014 | Safety requirements for lithium-ion batteries and battery packs for portable electronic products |
| UL 1642:2012 | Lithium Battery Standard |
| UL 2054: 2012 | Household and Commercial Battery Pack |
| UN38.3(2012) | Recommendations on the Transport of Dangerous Goods - Manual of Tests and Standards Part 3 |
| IEC62133-2012 | Safety requirements for batteries and battery packs containing alkaline or non acidic electrolytes |
| Adjustable range of test pressure | 10Kpa to 101.3Kpa |
| Control method | PLC touch screen human-machine interface control+computer remote control (dual control mode) |
| Inner box material | SUS304 # stainless steel plate+bottom square tube channel steel reinforcement, stainless steel plate thickness 10mm |
| Load bearing capacity of test box bottom plate | ≥2200kG |
| outer box material | Cold rolled steel plate paint treatment |
| Experimental space size | W2000XD3000XH2000mm |
| Equipment outer box size | W2500XD4000XH3000mm (excluding controller protrusion) |
| Load bearing capacity of turnover platform | More than 2000kg |
| Size of turnover table top | W1500XD2000XH200 mm |
| Box door | Single door, box door with locking device |
| Box door lock | The retaining component is a strong cold storage door handle, equipped with a locking device, with sufficient rigidity and strength, using strong explosion-proof hinges, with a quantity of 4 |
| Sealing of box door system | Using Toshiba high-purity silicone rubber raw materials, the sealing effect is good |
| lighting system | There are 2 LED explosion-proof lighting fixtures installed on the top of the laboratory, which can prevent dust, explosion, and fire. The explosion-proof lamp housing is easy to replace |
| Vacuum degree display | Digital vacuum degree control meter, with intuitive and easy to set digital display mode |
| Timing display and control | PLC human-machine interface display and control time |
| vacuum pumping time | Reduce from 101.3kPa to 11.3kPa within 5 minutes |
| Vacuum resolution and control accuracy | 0.1Kpa |
| Pressure relief method | There are two ways to relieve pressure: manual and automatic |
| vacuum pump | Taiwan Crown series vacuum pump |
| Recording function |
Sampling frequency up to 1Hz, capable of recording detection and setting test curves, data, operation records, alarm records within 12 months, printable
|
| Pressure relief protection | When the gas pressure inside the box increases, the pressure sensors near the pressure relief port can quickly open the top pressure relief port to discharge gas and heat, reducing the pressure inside the box |
| air source | 0.8~1MPA |
| Use power supply | AC380V 50HZ |
| Power | 15KW |
Learn moreBattery pack simulation high-altitude test boxPlease consult customer service for performance parameters, http://www.gbtest.net/