Handheld digital film thickness gauge Main technical specifications: Measurement range: (0-25) mm Resolution: 0.001mm Power supply: Silver oxide batte
Product details
brand other brands Origin category domestic
Application field printing and packaging

BMH-J3Handheld digital thin film thickness gauge

Instruction Manual

  • Handheld digital thin film thickness gaugeMain technical indicators:

Measurement range:(0-25mm

Resolution:0.001mm

Power supply: Silver oxide batterySR44

Operating Temperature0~+40℃

Storage and transportation temperature: -20~+70 ℃

Relative humidity: ≤ 80%

  • Main functions:

data output

Zero at any position

Conversion from metric to imperial system

Auto power-off

  • Instructions for Use:
  1. At the data output interface of position (1), data can be output,

The interface is a capacitive grid digital measuring tool interface.

  1. Press the ZERO key (3) to reset to zero at any position.
  2. The mm/in button (4) can be used to convert between metric and imperial systems.
  3. Automatically power off if not measured within ten minutes.
  4. Battery replacement method: Use a wrench with a round back as a wrench to unscrew the battery cover,

Replace the battery with a new one (with the positive electrode facing outward), and then screw on the battery cover.

  • Precautions for using digital thin film thickness gauge:
  1. Before use and calibration, please verify the zero position.
  2. Electronic digital display components must prevent oil, water, or other liquids from entering.
  3. The micrometer screw and measuring surface must be cleaned regularly and coated with anti rust oil after use.
  4. The digital thin film thickness gauge is a precision measuring tool and should be prevented from impact to avoid affecting accuracy.
  5. The movement distance of the differential cylinder should be within the range of the scale sleeve's markings, and the deviation from the markings should not exceed 0.5mm.
  6. When the thin film thickness gauge is not used for a long time, the battery should be removed.
  • Troubleshooting method for digital thin film thickness gauge:

Fault phenomenon

reason

exclusion method

Digital flashing display

Low battery voltage

Replace the battery with a new one

do not display

  1. In automatic power-off state
  2. Voltage below1.1v
  3. Poor battery contact
  1. Rotate the differential cylinder to restore display
  2. Replace the battery with a new one
  3. Make it in good contact

The monitor cannot count

Accidental circuit malfunction

Remove the battery, etc30Reinstall in seconds