BINDER KBF720 Constant Temperature and Humidity Chamber
1、 Product Introduction:
1. Temperature range: 0 ℃ to 70 ℃
2. Humidity range: 10% relative humidity to 80% relative humidity
3 APT.lineTMPreheating chamber technology
4. Using capacitive humidity sensors and steam humidification for humidity regulation
5. Intuitive touch screen controller using interval programming and real-time programming
6. Internal data logger and open format measurement values can be read through USB
7. Equipment self-test is used for comprehensive status analysis
8. The sealed inner door is made of safety glass (ESG)
9. Prevent corrosion of glass through a special TIMELESS coating.
10. The interior is entirely made of stainless steel
11. A bracket made of stainless steel
12. Access hole with silicone plug 30mm, left
13. 4 stable casters, 2 with brakes, starting from 240 liters
14. Independently adjustable temperature safety device Class 3.1 (DIN 12880), equipped with optical and acoustic temperature alarms
15. Computer interface: Ethernet
16. Door heating
2、 Technical parameters:
| PRODUCT MODEL | KBF720-230V1 | KBF720UL-240V1 |
| Choose a model | standard | standard |
| Item code | 9020-0324 | 9020-0325 |
| measure | ||
| Internal volume [L] | 700 | 700 |
| Net weight of equipment (unloaded) [kg] | 309 | 309 |
| Maximum total load [kg] | 150 | 150 |
| Maximum load per partition [kg] | 45 | 45 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 1250 | 1250 |
| Net weight at height [mm] | 1925 | 1925 |
| Depth net weight [mm] | 890 | 890 |
| Wall to side distance [mm] | 300 | 300 |
| Wall distance from rear [mm] | 100 | 100 |
| Internal dimensions | ||
| Width [mm] | 973 | 973 |
| Height [mm] | 1250 | 1250 |
| Depth [mm] | 576 | 576 |
| The number of doors | ||
| Inner door | 2 | 2 |
| Outer door | 2 | 2 |
| Data related to the environment | ||
| Energy consumption at 40 ℃ and 75% relative humidity [Wh/h] | 620 | 620 |
| Sound pressure level [dB (A)] | 53 | 53 |
| firmware | ||
| Number of partitions (hours/maximum) | 2/15 | 2/15 |
| Temperature performance data | ||
| Temperature range [℃] | 0…70 | 0…70 |
| Temperature deviation at 40 ℃ [± K] | 0.2 | 0.2 |
| Temperature fluctuation [± K] | 0.1 | 0.1 |
| The maximum thermal compensation is at 40 ℃ [W] | 600 | 600 |
| Climate performance data | ||
| Temperature range [℃] | 10…70 | 10…70 |
| Temperature deviation at 40 ℃ and 75% relative humidity [± K] | 0.2 | 0.2 |
| Temperature deviation at 25 ℃ and 60% relative humidity [± K] | 0.2 | 0.2 |
| Temperature fluctuation at 40 ℃ and 75% relative humidity [± K] | 0.1 | 0.1 |
| Temperature fluctuation at 25 ℃ and 60% relative humidity [± K] | 0.1 | 0.1 |
| Humidity range [% relative humidity] | 10…80 | 10…80 |
| Humidity fluctuation at 25 ℃ and 60% relative humidity | 1.5 ± relative humidity | 1.5 ± relative humidity |
| Humidity fluctuation at 40 ℃ and 75% relative humidity | 1.5 ± relative humidity | 1.5 ± relative humidity |
| The humidity recovery time after opening the door in 30 seconds is at 40 ℃ and 75% relative humidity [min] | 17 | 17 |
| The humidity recovery time after opening the door in 30 seconds is at 25 ℃ and 60% relative humidity [min] | 16 | 16 |
| Electric data | ||
| Rated voltage [V] | 200…230 | 200…240 |
| Power frequency [Hz] | 50/60 | 50/60 |
| Rated power [kW] | 3.1 | 3.1 |
| Equipment safety device [A] | 16 | 16 |
| Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.