BINDER FP240 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃
2 APT.line ™ Preheating chamber technology
3. Adjustable fan speed
4. Adjustable exhaust valve
5. Controller using interval programming and real-time programming
6. 2 chrome plated sockets
7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
8. Computer interface: RS 422
2、 Technical parameters:
PRODUCT MODEL | FP240-230V1 | FP240UL-208V1 |
Choose a model | standard | standard |
Item code | 9010-0263 | 9010-0264 |
measure | ||
Internal volume [L] | 240 | 240 |
Net weight of equipment (unloaded) [kg] | 96 | 96 |
Maximum total load [kg] | 70 | 70 |
Maximum load per partition [kg] | 30 | 30 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 1035 | 1035 |
Net weight at height [mm] | 825 | 825 |
Depth net weight [mm] | 745 | 745 |
Wall distance from rear [mm] | 160 | 160 |
Wall to side distance [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 800 | 800 |
Height [mm] | 600 | 600 |
Depth [mm] | 510 | 510 |
Data related to the environment | ||
Energy consumption at 150 ℃ [Wh/h] | 850 | 850 |
firmware | ||
Number of partitions (hours/maximum) | 2/7 | 2/7 |
Temperature performance data | ||
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] | 300 | 300 |
Temperature deviation is 150 ℃ [± K] | 2 | 2 |
Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
Heating time to 150 ℃ [min] | 27 | 27 |
The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 10 | 10 |
Ventilation data | ||
Air exchange rate at 150 ℃ [x/h] | 20 | 20 |
Electric data | ||
Rated voltage [V] | 230 | 208 |
Power frequency [Hz] | 50/60 | 60 |
Rated power [kW] | 2.7 | 2.7 |
Equipment safety device [A] | 16 | 3 x 16 |
Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.