AvaSpec-ULS2048x64-EVO
In addition to the refrigerated spectrometer AvaSpec-ULS2048x64TEC-EVO with a low-noise detector, Avantes also offers a more affordable non refrigerated spectrometer AvaSpec-ULS2048x64-EVO. With its standard 2048x64 back illuminated detector, this spectrometer is highly suitable for applications with low sensitivity requirements in the UV and NIR bands. For applications that require an integration time of less than 2 seconds, refrigeration is usually not necessary. This non cooled spectrometer AvaSpec-ULS2048x64-EVO has a high UV response and a 0.9 mm high detector, making it very suitable for application in the DOAS field.
The options include a second-order diffraction filter and an inflation port for deep ultraviolet measurement. AvaSpec-ULS2048x64-EVO has multiple slit sizes and gratings, and can be configured with SMA or FC/PC fiber optic connectors. AvaSpec-ULS2048x64-EVO uses AS-7010 circuit board, providing USB3 (10 times faster than USB2), Gigabit Ethernet, and better signal processing capabilities. By connecting to a PC via USB3 or Ethernet, the data transfer speed can reach 2ms/scan. The instrument comes with AvaSoft Basic software, USB cable, and operation manual.
technical data
optical table | ULS, symmetrical Czerny Turner optical path, focal length 75 mm |
wavelength range | 200-1160nm |
resolution | 0.09-20 nm, depending on spectrometer configuration |
stray | <1%, depending on the grating |
sensitivity | 650,000(Count/µ W, integration time per millisecond) |
detector | Back illuminated CCD detector, 2048x64 pixels (height: 0.89 nm) |
Signal to Noise Ratio | 450:1 |
ad converter | 16 bit, 1.33 MHz |
pointstime | 2.4 milliseconds -25 seconds |
communication interface | USB 3.0 high-speed, 5 Gbps, Gigabit Ethernet 1 Gbps |
Sample Rate | 2.44 ms/scan |
Readout Noise | 7.5 cnt RMS |
Data transmission speed |
2.44 ms/scan(USB3) |
Digital I/O interface | HD-26 connector, 2 analog inputs, 2 analog outputs, 13 digital bidirectional, trigger, sync, strobe, laser |
Power requirements | Default USB3 power supply, 885 mAOr 12 VDC, 420 mA |
Size, weight | 175 x 127 x 44.5 millimeters (single channel), 1180 grams |
Grating selection table
Application scope | Available wavelengths | Spectral range (nm) | Number of grating lines/mm | Sparkling wavelength (nm) | Grating model |
UV/VIS/NIR | 200-1160** | 960** | 300 | 300 | UA |
UV/VIS/NIR | 200-1100** | 900** | 300 | 300/1000 | UNA-DB |
UV/VIS | 200-850 | 520 | 600 | 300 | UB |
UV | 200-750 | 250-220* | 1200 | 250 | UC |
UV | 200-650 | 165-145* | 1800 | UV | UD |
UV | 200-580 | 115-70* | 2400 | UV | UE |
UV | 200-400 | 70-45* | 3600 | UV | UF |
UV/VIS | 250-850 | 520 | 600 | 400 | BB |
VIS/NIR | 300-1160** | 860** | 300 | 500 | VA |
VIS | 360-1000 | 500 | 600 | 500 | VB |
VIS | 300-800 | 250-200* | 1200 | 500 | VC |
VIS | 350-750 | 145-100* | 1800 | 500 | VD |
VIS | 350-640 | 75-50* | 2400 | VIS | VE |
NIR | 500-1050 | 500 | 600 | 750 | NB |
NIR | 500-1050 | 220-150* | 1200 | 750 | NC |
NIR | 600-1160 | 350-300 | 830 | 800 | SI |
NIR | 600-1160** | 560** | 300 | 1000 | IA |
NIR | 600-1160 | 500 | 600 | 1000 | IB |
*Depends on the starting wavelength of the grating; The higher the wavelength, the greater the dispersion, and the smaller the selection range.
**Please note that not all 2048 pixels will be used for the available range
Resolution Table of Spectrometer
Slit width (μ m) | ||||||
Grating (lines/mm) | 10 | 25 | 50 | 100 | 200 | 500 |
300 | 1.40 | 1.50 | 2.5 | 4.8 | 9.2 | 21.3 |
600 | 0.70-0.80* | 0.75-0.85* | 1.2 | 2.4 | 4.6 | 10.8 |
830 | 0.42-0.48* | 0.50-0.58* | 0.93 | 1.7 | 3.4 | 8.5 |
1200 | 0.25-0.31* | 0.37-0.43* | 0.52-0.66* | 1.1 | 2.3 | 5.4 |
1800 | 0.17-0.21* | 0.26-0.32* | 0.34-0.42* | 0.8 | 1.6 | 3.6 |
2400 | 0.12-0.18* | 0.18-0.24* | 0.26-0.34* | 0.44-0.64* | 1.1 | 2.7 |
3600 | 0.09-0.12* | 0.11-0.15* | 0.19 | 0.4 | 0.8 | 1.8 |
*Depends on the starting wavelength of the grating; The higher the wavelength, the greater the dispersion, and the smaller the selection range.
ordering information
AvaSpec-ULS2048x64-EVO | 75mm AvaBench, 2048x64 pixel backlit CCD detector, USB powered, high-speed USB3 and ETH interface. AvaSoft Basic, USB interface cable. Specify grating, wavelength range, and options. |
PS-12V/1.0A | The external power required to run in ETH mode. |