AvaSolar series spectroradiometer
The AvaSolar series spectroradiometer is specifically designed for spectral matching testing of solar simulators, and is suitable for measuring the spectral characteristics of continuous and pulsed (single flicker or high-frequency short pulse width) solar simulators. It can achieve high-speed measurement of 450 spectra per second and capture the spectral intensity at different time points within a pulse cycle. In addition, using non-linear calibration techniques, the response consistency is close to 100%. The spectral radiance provided by the AvaSolar series spectroradiometer can be traced back to the standard light source of the Chinese Academy of Metrology. The uncertainty of the calibration result is U=8.0% (k=2), and a calibration certificate from the Chinese Academy of Metrology is provided.
technical data
model |
AvaSolar-1-Pro |
AvaSolar-2-Pro |
AvaSolar-2-Compact |
AvaSolar-3 |
AvaSolar-HH |
spectral range |
300-1100 nm |
200-1700 nm |
200-1700 nm |
300-2500 nm |
300-1100 nm |
spectral resolution |
2.5 nm @ 300-1100 nm |
2.5 nm @ 200-1100 nm 5.7 nm @ 1050-1700 nm |
2.5 nm @ 200-1100 nm 8 nm @ 1050-1700 nm |
2.5 nm @ 300-1100 nm 15 nm @ 1050-2500 nm |
2.5 nm @ 300-1100 nm |
Spectral sampling interval |
0.6 nm @ 300-1100 nm |
0.6 nm @ 200-1100 nm; 3 nm @ 1050-1700 nm |
0.6 nm @ 200-1100 nm 3.3 nm @ 1050-1700 nm |
0.6 nm @ 300-1100 nm; 6 nm @ 1050-2500 nm |
0.6 nm @ 300-1100 nm |
Spectral accuracy |
± 0.3 nm @ 300-1100 nm |
±0.3nm @ 200-1100 nm; ±1.5nm @ 1050-1700 nm |
± 0.3 nm @ 200-1100 nm ±1.5 nm @ 1050-1700 nm |
± 0.3 nm @ 300-1100 nm ±3 nm @ 1050-1700 nm |
± 0.3 nm @ 300-1100 nm |
detector |
Back illuminated CCD detector, 2048 pixels, Pixel size 14x500um |
Back illuminated CCD detector, 2048 pixels, pixel size 14x500um 256 pixel InGaAs array, TE cooling |
Back illuminated CCD detector, 2048 pixels, pixel size 14x500um 256 pixel InGaAs array |
Back illuminated CCD detector, 2048 pixels, pixel size 14x500um 256 pixel InGaAs array, TE cooling |
Back illuminated CCD detector, 2048 pixels, Pixel size 14x500um |
Signal to noise ratio |
525:1 |
525:1@ 200-1100 nm 5000:1 @1050-1700 nm |
525:1@ 200-1100 nm 5000:1 @1050-1700 nm(LN) |
525:1@ 300-1100 nm 4000:1 @1050-2500 nm |
525:1 |
Integral time (exposure time) |
2us – 20s |
||||
Wavelength Repeatability |
± 0.1 nm @ ± 10 ° C temperature variation |
||||
Hardware Spectral Average |
Up to 100000 times |
||||
Response linearity |
Better than 99% |
ordering information
model |
AvaSolar-1-Pro |
AvaSolar-2-Pro |
AvaSolar-2-Compact |
AvaSolar-3 |
AvaSolar-HH |
|
Size/Weight |
175x110x44mm/0.8kg |
310x270x135mm/5.8kg |
290x170x120mm / 3.5kg |
310x270x135mm/5.8kg |
215x161x108mm /2kg |