Detailed description of Tektronix TDS3012C digital fluorescence oscilloscope:
Main performance indicators:
100MHz, 300MHz, and 500MHz bandwidths, with 2 or 4 channels
Real time sampling rate of up to 5GS/s on all channels, with a recording length of 3600wfms/s and a continuous waveform capture rate of 10k on all channels
Advanced trigger kit

TDS3000C series digital fluorescence oscilloscope
TDS3012C-100 MHz, 2 channels, 1.25 GS/s.
TDS3014C-100 MHz, 4 channels, 1.25 GS/s.
TDS3032C - 300 MHz, 2 channels, 2.5 GS/s.
TDS3034C - 300 MHz, 4 channels, 2.5 GS/s.
TDS3052C-500 MHz, 2 channels, 5 GS/s.
TDS3054C-500 MHz, 4 channels, 5 GS/s.
Easy to use features:
Front panel USB host port for easy storage and transfer of measurement data
25 automatic measurement functions
Standard FFT
Multilingual user interface
WaveAlert © Automatic detection of abnormal waveforms using TekProbe © The interface supports active probes, differential probes, and current probes, with automatic calibration and unit determination
Portable design:
Lightweight design (only 7 pounds/3.2 kilograms), easy to carry
Optional built-in battery, can work for up to three hours without power supply
Specialized analysis application module:
Advanced Analysis Module
Extreme testing module
Communication template testing module
Expand video module
601 Serial Digital Video Module
Provide the performance you need at an economical price: The TDS3000C series digital fluorescence oscilloscope (DPO) provides the performance you need at an economical price. TDS3000C provides a bandwidth of 100 MHz-500 MHz and a sampling rate of up to 5 GS/s, which can accurately represent signals.
DPO can provide a more comprehensive view of complex signals:
To solve a problem, it is necessary to first see the problem. The TDS3000C series offers a continuous waveform capture rate of 3600 wfms/s and real-time brightness levels, allowing you to see and solve problems.
The fast continuous waveform capture rate can quickly reveal the characteristics of the problem, and then apply advanced trigger isolation problems, saving a lot of time.
The real-time brightness level highlights details related to the signal activity history, making it easier to understand the waveform characteristics captured. Unlike other similar oscilloscopes, it can still retain historical data even after the acquisition stops.
Quickly debug and verify signals using DRT sampling technology and Sin (x)/x interpolation function:
The TDS3000C series combines unique digital real-time (DRT) sampling technology with sin (x)/x interpolation function, allowing for precise calibration of various signal types on all channels simultaneously. Unlike other similar oscilloscopes, the sampling rate of TDS3000C series remains unchanged when additional channels are opened. This sampling technique can capture high-frequency information that may be missed by other similar oscilloscopes, such as spikes and abnormal edges. Sin (x)/x interpolation ensures accurate reconstruction of each waveform.
Application scope:
Digital design, debugging, and testing
Video installation and service
Power supply design
Training
Communication template testing
manufacture test
Universal Workbench Testing
TDS3000C series electrical characteristics
TDS3012C TDS3014C TDS3032C TDS3034C TDS3052C TDS3054C
Bandwidth 100 MHz 100 MHz 300 MHz 300 MHz 500 MHz 500 MHz
Calculated rise time
(Typical value) 3.5 ns 3.5 ns 1.2 ns 1.2 ns 0.7 ns 0.7 ns
Number of channels 2 4 2 4 2 4
External trigger input includes all models
Sampling rate on each channel 1.25 GS/s 1.25 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s 5 GS/s
Maximum record length 10K sample points
Vertical resolution of 9 bits
Vertical sensitivity, 1 M Ω 1 mV/div -10 V/div
Vertical sensitivity, 50 Ω
1 mV/div - 1 V/div
Input coupling AC, DC, GND
When the input impedance is connected in parallel with 13 pF or 50 Ω terminals, it is 1 M Ω
DC gain accuracy ± 2%
Maximum input voltage (1 M Ω) 150 VRMS, peak value ≤ 400 V
Maximum input voltage (50 Ω) 5 VRMS, peak value ≤ 30 V
Positioning range ± 5 div
Bandwidth limit 20 MHz 20 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz 20 MHz, 150 MHz
Time base range 4 ns-10 s 4 ns-10 s 2 ns-10 s 2 ns-10 s 1 ns-10 s 1 ns-10 s
Time base accuracy ± 20 ppm at any 1 ms time interval
Input/output interface
Ethernet port RJ-45 connector, supporting 10Base-T LAN
USB port front panel USB 2.0 host port, supports USB flash memory
Comprehensive call/answer modes, settings, and measurements for GPIB ports (optional TDS3GV communication module)
RS-232-C port DB-9 male connector, with comprehensive call/answer modes; Control all modes, settings, and measurements
The highest baud rate is 38400 (optional TDS3GV communication module)
VGA video port DB-15 female connector, monitor output, directly displays information on large monitors equipped with VGA (TDS3GV communication module optional)
External trigger input BNC connector, input impedance>1 M Ω, parallel current 17 pF; maximum input voltage 150 VRMS
Collection mode:
DPO - Capture and display subtle patterns in composite waveforms, random events, and actual signal behavior. DPO can display, store, and analyze the three-dimensional information of signals in real time, including amplitude, time, and amplitude distribution in each period.
Peak detection - captures high-frequency and random spikes. Can capture burrs as narrow as 1ns.
WaveAlert - Monitor input signals on all channels and alert users when any waveform deviates from the captured normal waveform.
Sampling - Data sampling only.
Average - Calculate an average of 2 to 512 (optional) captured waveform data.
Envelope - the maximum/minimum value obtained on one or more captured signals.
Single Sequence - Use the Single Sequence button to capture one single triggered capture sequence at a time.
Trigger system:
Main triggering methods - automatic triggering (supporting 40ms/grid and slower scrolling modes), normal triggering, and single sequence triggering.
B trigger - triggered after time or event.