Adjustment of microscope illumination system
This article is organized and provided by the inverted microscope. In order to provide uniform and sufficient illumination for the field of view of th
Product details

This article is organized and provided by an inverted microscope

In order to provide uniform and sufficient illumination for the field of view of the microscope, the illumination light path system must be adjusted properly during the initial installation and debugging of the microscope. This is an important means and basic requirement for the correct use of the microscope and obtaining accurate results. In addition, correctly mastering the adjustment of the lighting path system is a necessary step after replacing the light source bulb during the use of the microscope, and it is also a necessary means to periodically check the performance of the microscope during daily use. The adjustment of the microscope illumination light path system mainly includes the following4Item content:

  1. Preliminary adjustment of lighting source lamp room outside the microscope
Firstly, open the outer shell of the lamp chamber and insert the halogen bulb into the socket using the spring clip. During installation, avoid direct contact between fingers and the bulb (which can be separated by a soft cloth or paper) to prevent fingerprints and other dirt from leaving on the bulb, which may affect its service life.
Place the lamp chamber on the table, turn on the power, and use a dedicated screwdriver to adjust the focusing knob hole of the lamp (marked with "←→"), so that the filament is projected onto the table1-2mOn the outer wall, adjust the filament imaging to clear; Then adjust the high and low position of the lamp to adjust the filament hole (marked with "-") so that the filament position is appropriate; Adjust the screw holes (marked with "-") of the left and right positions of the lamp again to make the filament position suitable. 【 12-2 】

Technical extension, welcome to log in:
&bid=26



Shanghai Huxing Optical Instrument Co., LtdMain production and salesmetallurgical microscopepolarization microscopestereomicroscopeTesting microscopeVickers hardness testerWaiting for optical equipment.Sales department hotline:

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!