VIP member
9350&9550AiroTrakTM portable laser particle counter
TSI's 9350 and 9550 portable particle counters have excellent performance. The flow rate is 50 liters per minute, and the range of small particle size
Product details
AeroTrakTM Portable Laser Particle Counter Product Overview:
AeroTrakTMPortable laser particle counter
Type 9350&9550
TSI's 9350 and 9550 portable particle counters have excellent performance. The flow rate is 50 liters per minute, and the smallest particle size range that can be detected is 0.3 (9350) and 0.5 (9550) m. They fully comply with the ISO14644-1 standard. The sampling volume is 1 cubic meter every 20 minutes. The 9350 and 9550 counters are compatible with TSI's ventilation probe, which can measure air flow rate/temperature/humidity, providing a complete solution for demonstration.
The 9350 and 9550 particle counters can generate pass/non conformance reports based on ISO14644-1EU GMP Annex 1 and FS209E. It can not only be used as an independent particle counter, but also compatible with the entire monitoring system of FMS 5.
The 9350 and 9550 particle counters strictly comply with the ISO 21501-4 protocol. It is validated using standard particle measurement methods, which involve using PSL balls from the National Institute of Standards and Technology in the United States to pass through TSI * sieves and coagulation particle counters. Based on its two-year warranty and TSI's high-quality long-term good reputation, there is currently no particle counter of the same type on the market.
Features and advantages
AeroTrakTMThe particle size range of 9350 particle counter is 0.3 to 25 µ m
AeroTrakTMThe particle size range of 9550 particle counter is 0.5 to 25 µ m
Zui can simultaneously measure data from six channels
The flow rate is 50 liters per minute
Fully compliant with ISO 21501-4 requirements
10000 data storage capacity, 999 locations
Ethernet, USB output
Industrial high noise alarm
Simple icon touch screen operation
Provide reports ISO14644-1, EU GMP Annex 1, and FS209E
Allow storage and playback of specific tests
Zui High Display Three Environmental Parameters
Stainless steel accessories
Lightweight
It can be used independently or compatible with the entire monitoring system
Can be validated for cleanrooms that meet ISO Level 2 standards
specifications
Type 9350&9550
AEROTRAKTMPortable laser particle counter
| characteristic | 9350 & 9550 | | ||
| particle size range | 9350-02 | 9550-02 | ||
| | 9350:0.3-25 ? m | 9550:0.5-25 ? m | ||
| Particle size channel | 9350-02:0.3,0.5,1.0, | 9550-02:0.5,0.7, 1.0, | ||
| | 3.0, 5.0, 10.0 ? m | 3.0, 5.0, 10.0 ? m | ||
| | (Additional channels are optional) | (Additional channels are optional) | ||
| Particle size resolution | 9350-02: At 0.5? m <15% | 9550-02: At 0.5? m <15% | ||
| | (Compliant with ISO 21501-4) | (Compliant with ISO 21501-4) | ||
| counting efficiency | 9350-02:0.3 ? 50% at m, | 9550-02:0.5 ? 50% at m, | ||
| | Greater than 0.45? 100% at m | Greater than 0.75? 100% at m | ||
| | (Compliant with ISO 21501-4 and JIS) | (Compliant with ISO 21501-4 and JIS) | ||
| Concentration limit | Maximum loss of 5% at 250000 units/cubic foot | |||
| Laser source | Long life laser diode | | ||
| Zero adjustment | <1 every 5 minutes (compliant with ISO 21501-4 and JIS B9921) | |||
| internet traffic | 50 liters/minute ± 5% (compliant with ISO 21501-4 and JIS B9921) | |||
| flow control | Electronic automatic closed loop (flow control technology) | |||
| External pipeline | Up to 12 meters in length | | ||
| sampling mode | Manual, automatic, buzzing. Accumulated/difference; Counting or concentration | |||
| sample time | 1 second to 99 hours | | ||
| sampling frequency | 1 to 9999 cyclic or continuous type | |||
| exhaust | Built in HEPA filter membrane | | ||
| Power source | Built in pump | | ||
| communication mode | Modbus®RTU via Ethernet or USB interface | |||
| Data storage capacity | 10000 data records, including date, time, and six others | |||
| | Channel, flow rate, sample ID, sampling flow rate, via USB | |||
| | Transfer data or use TraxProTMSoftware Download | |||
| Audible warning | Built in, greater than 85dB at 1 meter (adjustable) | |||
| status indicator | Low battery, data flow, laser | | ||
| Alarm Output | Dry contact point, closed upon alarm | | ||
| Alarm limit | Different channels programmable | | ||
| display | VGA 14.5-cm touch screen | | ||
| language | Simplified Chinese, English, Japanese, German, French, Spanish | |||
| | Dental and Italian. | | ||
characteristic | 9350 & 9550 | | |||
software | Compatible with TrakProTMLite and FMS 5 software | | |||
printer | Built in thermal printer (optional without printer) | | |||
print out | Available in all languages | | |||
Environmental sensor interface | Support TSI probe model: 960962964966 (air) | | |||
| Flow rate/temperature/humidity) and temperature/humidity probes | | |||
Unit Number | Configurable IP address | | |||
secure | Second level password protection | | |||
Local ID | Up to 999 locations; Each 16 character length | | |||
report | Provide ISO 14644-1, EU GMP and | | |||
| Approval/Non Compliance Report for FS209E | | |||
check | Traceable NIST TSI Verification System | | |||
Verification frequency | Recommended at least once a year | | |||
Volume (height x length x width) 24.1cmx23cm x 23.2cm x 23.2cm (excluding constant velocity sampling port) | | ||||
weight | 6.44 kilograms (including two batteries) | | |||
quantity of electricity | 110 to 240VAC, regular power supply | | |||
battery | Detachable rechargeable lithium battery, 2 battery slots | | |||
BATTERY LIFE | Can be used continuously for more than 3 hours | | |||
charging time | 4 hours (idle time) | | |||
surface material | stainless steel | | |||
Operation range | 5-35℃, 20%-95%Non condensing | | |||
Storage range | 0-50℃, 98%Non condensing | | |||
standard | ISO 21501-4,CE,JIS B9921 | | |||
warranty | Two years, providing extended warranty | | |||
Included Accessories | Quick Start Manual, CD-ROM Operating Manual, Power Supply | | |||
| Wire, battery, sampling port, constant velocity probe, tripod, tube | | |||
| Road, filter, printing paper, USB cable, and TrakProTM
| | |||
| Lite software | | |||
Options | Electronic filtering scanning probe, basic filtering scanning probe, | | |||
| TSI wind speed probe, temperature and humidity probe, backup battery, dual | | |||
| External charger, constant velocity probe, constant velocity sampling port, sampling | | |||
| Sample tube, portable case | |
Online inquiry