[Functional Features]
❖All in Chinese7.0Inch color touch screen
❖Follow national standards and use potassium dichromate method
❖Automatic endpoint recognition, dynamic titration with equal volume titration
❖Adopting a fully automatic robotic arm for titration, it is more convenient, fast, and accurate
❖Automatic stirring during titration results in a more uniform and clear color development of the solution
❖mvMeasurement electrode interface, hostUSBThe interface can realize data upload
Technical indicators
●Measurement range:3-750mg/L(When it exceeds the upper limit of the measurement, the measurement can be diluted)
● AccurateDegree:relative error<2%
●Automatic burette:1Built in(Number of titration controls used)
●Pipette pump:3Built in (for adding auxiliary reagents)
●position Number:7Positional system(6A sample grade,1A cleaning position
●Cleaning method: single cleaning
●Titration cup capacity:150-200ml
●Mixing method: non-contact magnetic stirringElectrode interface type:MVMeasurement electrode interface/hostUSBInterface
●Measurement time: ≤15Minutes per sample
●Product size:1000X430X700mm
●weight/Power:39kg/200W
6B-8CtypeCODbackflowDissolving instrument:
[Functional Features]
※ Dual fan convection cooling, independently controllable, with significant cooling effect
※ Large screen LCD Chinese display, adjustable time and power, users can modify, set, and save on their own
Using microcrystalline heating panel and infrared heating method, the heating speed is fast and uniform, and the service life is longer
After the digestion is completed, the instrument will automatically sound a warning and stop heating work
※ Water saving, low thermal radiation, and small space occupation
※ Easy to operate, can effectively reduce costs and improve work efficiency
Technical indicators
Measurement range: 4-700mg/L, 700-1000mg/L (requires diluted water sample)
Measurement time:≤ 150 minutes (can be set according to user needs)
Measurement error: Potassium hydrogen phthalate standard solution (500mg/L), relative standard deviation not exceeding 5.0%, industrial organic wastewater (500mg/L), relative standard deviation not exceeding 8.0%
※ Environmental temperature:- 5~40℃
※ Display mode: LCD display
※ Sample size for digestion: 8 per batch
※ Dissolution time:Users can set their own time according to their needs,1-999 minutes
※Fan settings: automatic/Manual
Heating power: 1200W (AC 220V ± 22V, 50HZ)