Product Introduction
The 3330 Optical Particle Size Spectrometer (OPS) is a lightweight and portable device that utilizes single particle counting technology to provide fast and accurate detection of particle concentration and size distribution. OPS, with over 40 years of experience in aerosol instrument design, utilizes a 120 ° light collection optical system and sophisticated electronic processing equipment to obtain accurate and high-quality data. Strict factory calibration standards ensure testing accuracy. The 3330 optical particle size spectrometer can be used independently or placed in TSI's waterproof and environmentally friendly box.
application area
Quality optical instruments are widely used in various application fields due to their ease of use, fast detection time, robustness, and reliability. Below are some common applications listed.
Filter testing (i.e. ASHRAE 52.2)
Indoor air quality
Workplace monitoring
Outdoor environmental monitoring
Industrial measurement
Emission monitoring and control
main features
Particle size resolution less than 5% at 0.5 microns
User adjustable particle size channel
Particle size range for detection: 0.3-10 microns, up to 16 channels
Particle size sensing range: 0.5 to 25 mm
Detection concentration range: 0-3000/cm3
Color touch screen, intuitive user interface
Fully meets the requirements of ISO 21501-01/04
The refractive index and density can simultaneously display the number concentration and mass concentration of particles
The sampling membrane of the collection can be used for weighing measurement and chemical analysis
The battery level can provide 20 hours of operation
Up to 30000 data can be stored
Record the temperature and pressure of each sample
Technical Parameter
Measurement principles |
120°Light scattering and membrane sampling |
Concentration limit |
highest3,000individual/cm3(3,000,000individual/L) |
mass concentration |
0.001-275,000mg/m3 |
Particle size of particulate matter |
|
Detecting particle size range |
0.3-10mm |
Particle size resolution |
0.5mmtime5%(Compliant with)ISO 21501-01/04) |
Particle size channel |
at most16Channel, user adjustable |
Time |
≥1sUser adjustable |
midnight |
Every minute<1individual |
internet traffic |
|
Sample |
1L/min±5%Accuracy (compliant)ISO 21501) |
Sheath gas |
1L/min |
environmental condition |
|
operate temperature |
0-45℃ |
Operating humidity |
0-96%, without condensation |
Storage temperature |
-20-60℃ |
Aerosol medium |
air |
Data Storage |
5MBBuilt in storage(30,000Sample) |
communication |
|
interface |
USBInterface, Ethernet, andUtray |
digital display |
5.7Inch color touch screen, drawing display |
Analog Output |
0-5Vperhaps4-20mA,User Selectable |
Alarm Output |
Visualization, buzzer, relay or switch quantity, user set alarm value |