The medium temperature blackbody radiation source is mainly used to calibrate radiation thermometers in the temperature range of (300-1100) ℃. It uses resistance wire heating and can be used in an oxidizing atmosphere. One end is open. In order to improve the effective emissivity, the isothermal blackbody cavity method is used, which involves winding furnace wires unevenly or uniformly winding both ends with auxiliary winding heating to produce a uniform temperate zone. The cavity is isothermal and forms an isothermal blackbody.
Including the metrology certificate of the National Defense Science and Technology Industry First Metrology Testing and Research Center, the technical indicators meet the requirements of current regulations/specifications.
Temperature range: (300-1100) ℃
Effective radiation surface: Ф40mm
Effective emissivity: 1 ± 0.002
Maximum allowable error: ± (0.2% t+1) ℃
Temperature stability: not exceeding 0.1% t/10min
Temperature resolution: 0.1 ℃
Temperature control: PID automatic control
Operating environment temperature: (5-40) ℃
Power supply: 220VAC ± 10%
Maximum power: 2kW
Weight: 20kg
Overall dimensions: 480mm × 280mm × 440mm
Requirements for installation supporting basic conditions:
Power socket: 220V 10A
Area requirement: Small in size, can be placed on a workbench