Product brief
YKT-03 oil laser particle counter is based on standards such as GB/T18854-2002- (ISO11171-1999). Developed using the principle of light blocking counting, it fully complies with the corresponding national military standards, national standards, and international standards. It can provide fast, accurate, reliable, and reproducible detection results and complete pollution monitoring and analysis reports, suitable for detecting particle pollution levels in hydraulic oil, lubricating oil, fire-resistant oil, insulation oil, and turbine oil. It can be widely used in aviation, aerospace, power, petroleum, chemical, transportation, ports, metallurgy, machinery, automobile manufacturing and other fields.
1. Adopting the counting principle of photoresist (shading) method developed by the International Hydraulic Standards Committee
2. High precision laser sensor with a wide testing range, stable performance, low noise, and high resolution
3. Adopting high-pressure injection pump sampling method, the sampling volume can be set by oneself, the injection speed is stable, and the sampling accuracy is high
4. The injection system combining positive and negative pressure can achieve sample degassing and is suitable for testing samples with different viscosities
5. Equipped with a built-in pressure sensor, it can directly set the pressure value and automatically determine the pressure inside the cabin to ensure safety
6. Built in air purification system ensures that testing is not contaminated
7. Built in SAE4059E SAE4059F、SAE4059F-CPC、NAS1638、ISO4406、GB/T14039、GJB-420A、GJB-420B、ГOCT17216-71、DL/T1096、SAE749D、JBT9737.1、QC/T29104-201、DL-T432-2018、
HH005-2018 and more than ten other commonly used standards support adding custom standard tests at will with no upper limit on the number of custom standards added, and the required standards can be set according to customer needs
8. Built in data analysis system, one test can provide all standard test data and pollution levels
9. Built in multiple calibration curves, compatible with all commonly used domestic and international standards, for single or multi-channel automatic calibration
10. The particle size can be set arbitrarily, with nearly 10000 built-in particle sizes for easy particle size analysis
11. Multiple sampling containers such as standard sampling bottles or cups can be used to meet the testing requirements of different industries
12. Optional viscosity, dielectric constant, moisture, and temperature sensor modules for multi parameter testing
13. Windows operating interface, built-in database, unlimited data storage, can be searched and queried using various methods such as testing standards, sample names, sample batch numbers, testing personnel, testing time, etc
14. Multi level account permission management, allowing for separate account permissions and supporting digital signatures
15. The test results can be printed using a micro or desktop printer, and can be exported in PDF format excelTables and other formats can be accessed throughUSB, WiFi and other transmission methods
1. Optical system: photoresist method (shading type)
2. Light source: semiconductor laser
3Measurement range:0.8um~600um (depending on different sensors)
4Measurement channel:64 channels (actual number of channels can be edited simultaneously), particle size can be set arbitrarily by the user
5Sensitivity:0.8um (ISO4402) or 3um (c) (GB/T18854, ISO11171)
6Resolution:<10%(GB/T18854,ISO11171)
7Repeatability:RSD<2%
8Test sample viscosity:≤ 400cSt (can be tested with heating or dilution methods when viscosity is too high)
9Sampling volume:0.2~6000ml, with an interval of 0.1ml
10Sampling speed:5 ~80mL/min
11Sampling volume accuracy: better than±0.5%
12. Pressure chamber: positive and negative pressure chamber device, realizing sample degassing and high viscosity sample detection
13Maximum vacuum in the pressure chamber:-0.08MPa
14Maximum positive pressure in the pressure chamber:0.8 MPa
15Overlap error limit:12000~40000 particles/mL
16temperature collectionT (optional):- 40~100℃Accuracy:±0.5℃
17Moisture (optional): Water activity collectionAW: 1~100% RH Accuracy: ± 1% RH
absolute water contentAH: 1~300ppm Accuracy: ± 10ppm
18Dielectric constant (optional):1~6
19Viscosity (optional):0.5-50cP accuracy: ± 2%(>10cP),±0.2cP(<10cP)
20Power Supply:110~245V AC, 50Hz, 70W
21. Dimensions: 340mm ╳ 310mm ╳ 730mm