uAccurate and rapid determination of COD, ammonia nitrogen, total phosphorus, and total nitrogen content values
uMulti parameter design, one instrument can complete the detection of multiple parameters, and the parameters can be combined as needed.
uAdopting advanced technologies such as imported light sources, narrowband interference, and software traps, it has high accuracy and good stability.
uAdopting a large screen LCD display, convenient and intuitive.
uOne click storage function, 1800 sets of data storage function.
uIt can store 80 standard working curves and can be self calibrated. Equipped with a small printer, measurement data can be printed in real-time (measurement numbers, measurement dates)
uThe digester is made of heat-resistant and corrosion-resistant materials, which are durable and long-lasting.
uIntelligent PID temperature control technology, uniform heating, fast heating time
uThe digester is suitable for the digestion of COD, total phosphorus, total nitrogen, total chromium and other projects.
uThe temperature of the digester can be flexibly set according to the user's sample requirements.
uAutomatic temperature control, anti overheating protection system, displaying current temperature, set temperature, and time.
product name |
Printed COD, ammonia nitrogen, total nitrogen analyzer (four parameters can be combined arbitrarily) |
||
PRODUCT MODEL |
YCN-3D |
||
detection range |
COD: 5-5000mg/L (segmented detection) Ammonia nitrogen: 0.02-60.00mg/L Total nitrogen: 0.05-100mg/L |
||
check the accuracy |
COD、 Total nitrogen: ± 5% F.S; Ammonia nitrogen: ± 5% F.S; |
||
repeatability |
Less than equal3% |
||
resolution |
0.001mg/L |
||
Optical stability |
0.002A/20min, imported cold light source |
||
Anti chlorine interference |
[Cl -]<1200mg/L has no effect (during COD determination) |
||
Light source lifespan |
Continuous use up to 100000 hours |
||
Dissolve quantity |
Can simultaneously digest 16 and 25 (optional) samples, with a digestion aperture of 16mm |
||
Dissolving temperature |
COD:165℃(or any setting between 0-200 ℃) When the timer ends, a buzzing prompt will sound Total nitrogen: 125 ℃ Ammonia nitrogen: No need for digestion, direct detection
|
||
digestion time |
COD: 15 minutes(or any setting from 0-9999min) When the timer ends, a buzzing prompt will sound Total phosphorus and total nitrogen: 30 minutes |
||
Temp Accuracy |
0.3℃ |
||
Printing function |
Small thermal printer, real-time data printing |
||
data transmission |
USB (data can be transferred to a computer) |
||
maximum power |
550W temperature automatic control, anti overheating protection system |
||
Product specifications |
|||
Product weight |
Host:< 3.5KG, disintegrator: < 7KG |
ambient temperature |
0-60℃ |
supply voltage |
220V,50HZ |
ambient humidity |
0-90% |
Host size |
242 * 320 * 160 (length * width * height) |
Dissolver size |
330 * 227 * 118 (length * width * height) |