YCA-1000 (AFG) fully automatic atomic absorption spectrophotometer
(Flame+Graphite Furnace) Fully Automatic Type
1、 Performance indicators
1Optical system
Wavelength Range |
190nm—900nmAutomatic adjustment |
Grating line density |
1800Line/mm |
Monochromator |
Aberration correctionCzerny-TurnerType monochromator |
Spectral bandwidth |
0.1、0.2、0.4、1.0、2.0nmMulti level automatic switching |
Wavelength accuracy |
±0.10nm |
Wavelength repeatability |
0.05nm |
Baseline drift |
static state0.002A/30mindynamiczero point zero zero six |
A/30min |
detector |
Wide spectral range photomultiplier tube |
resolving powerCan distinguish manganese lamps279.5nmand279.8nm20Double lines, and the energy of the valley between the two spectral lines should be less than |
2%
Light source system |
8Lamp holder |
Automatic switching of lamp holder |
Lamp power supply mode400Hz |
Square wave pulse |
Lamp current adjustment range2 — 10mA |
3average current
Atomization system | |
Flame analysis methoddetection limit |
0.003(Cu)μ |
g/ml |
burner100mm |
All titanium metal burner |
PrecisionRSD≤ |
0.5% |
Spray |
Metal sheathed high-efficiency glass spray |
Atomization chamber |
Corrosion resistant all plastic atomization chamber |
Security measures |
Equipped with multiple automatic safety protection functions: acetylene leak alarm, automatic shutdown system, etc | |
Flameless (graphite furnace) analysis methodDetection limit(Cd |
)<0.4*10ˉ |
12 g |
4Hydride generator atomic absorption method
Flame analysis method |
measuring methodair- |
Acetylene flame method |
Concentration calculation method |
Standard curve method, standard addition method, interpolation method. |
Repeated measurement times1~99 |
Next, automatically calculate the mean, content, standard deviation, and relative standard deviation. |
Result printingPrint test data or final analysis report. The results can be exported asExcel |
Worksheet. |
software environmentWindows XP/ Windows 7 |
5Operating system, professional Chinese software
Flameless (graphite furnace) analysis method |
measuring method |
Graphite furnace peak height method, graphite furnace peak area method |
Concentration calculation method |
Standard curve method, standard addition method, interpolation method. |
Background correction |
Deuterium lamp background correction greater than 60 times, self absorption buckle background greater than 60 times |
Repeated measurement times |
1-30 times, calculate the average and provide the standard deviation and relative standard deviation |
Result printingPrint test data or final analysis report. The results can be exported asword |
Worksheet. |
software environmentWindows XP/ Windows 7 |
Operating system, professional Chinese software
2、 Instrument parameters |
Power Supply |
AC 220V/50HZ |
power |
150VA |
Net weight of the instrument |
55kg |
Gross weight of the instrument |
80kg |
Instrument volume1110mm(Horizontal)×400mm(Depth) |
× 500mm |
(High)Packaging volume1350mm(Horizontal)×600mm |
(Depth) |
× 850mm(High) |
Working environment temperature |
10~35℃ |
Working environment humidity
1≤
280%
33、 Standard configuration
4. Host
Low noise oil-free air compressor. Hollow cathode lamp
Operating software
fourSource of procurement for core components of instrumentsPhotomultiplier tube: Hamamatsu Corporation, Japan
Five
2YCA-1000AFG
3Design features of fully automatic atomic absorption spectrophotometer
41. Total reflection optical system, full band chromatic aberration reduction, automatic wavelength scanning, and automatic peak searching.Equipped with deuterium lamp buckle background and self-priming buckle background..
5The uniquely designed all titanium burner has smooth combustion and fast thermal equilibrium speed.
.