XRF-100SPetroleum product sulfur content analyzer (XX-ray spectroscopy)
---XFluorescence sulfur analyzer
Product Features
The physical method of the instrument system adopts X-ray spectroscopy, which does not come into contact with or damage the sample during analysis, and does not require chemical reagents.
① Wide detection range: determination of sulfur content in petroleum and various petroleum products.
② Fast measurement speed: It only takes 30-60 seconds to accurately determine the sulfur content.
③ Easy to operate: The software runs on a PC with Windows 2000/Windows xp/Windows 7 installed, and all measurement work can be completed with just one click; Users can also set various measurement and calibration parameters according to their needs.
④ Complete historical records: According to user needs, the raw measurement data, measurement results, and measurement reports can be stored in the software's built-in database for long-term viewing, avoiding tedious measurement data maintenance work for users; The measurement results can also be exported in various forms such as spectra, Word documents, Excel spreadsheets, PDF documents, etc., and managed by users themselves.
⑤ Sample cup: using disposable sample cups, low cost, can avoid sample cross contamination; Sample production is fast and convenient, and not easily leaked; The single facial mask sample cup can be used to record the sample concentration, number, date, etc. on the back of the sample cup.
⑥ Low consumption of standard samples: only 4-10 milliliters are needed each time, greatly reducing the consumption of standard samples.
⑦ Easy instrument calibration: In addition to standard calibration, manual calibration of the standard curve is also used, which means that the values measured by the standard curve are re entered into the self checking system, and the instrument will automatically calibrate, reducing the consumption of standard samples.
⑧ Equipped with automatic gain correction: When the X-ray tube is used for a long time and the energy decreases, the instrument will automatically adjust the corresponding algorithm to ensure measurement accuracy, which actually extends the lifespan of the X-ray tube.
⑨ Automatic temperature and atmospheric pressure correction: The instrument can automatically adjust the internal system of the instrument based on the ambient temperature and atmospheric pressure, correcting the interference of the two changes on the sample. Users do not need to calculate correction factors to correct.
⑩ Automatic selection of standard curves: The instrument can store any number of standard curves, and users can choose the appropriate curve or select "automatic" to automatically select the curve that matches the measured sample. If the tested sample exceeds the range of all curve measurement points, the instrument will use the "extrapolation method" for measurement.
Pollution protection: An overflow protection device is added to the instrument sample detection chamber to prevent sample leakage during the detection process, which may cause pollution to the X-ray tube and detector.
Convenient connection: The instrument adopts a standard USB interface, which can be connected to any computer. High cost performance ratio, suitable for China's national conditions.
Instrument configuration: Instrument host, fluorescence sulfur measurement software, laptop (or desktop computer), inkjet or laser printer, etc.
give birth to article skill art instrument for drawing circles check
Compliant with standards |
GB-T 17040-2008 Petroleum and petroleum products - Determination of sulfur content - Energy dispersive X-ray fluorescence spectroscopy method GB-T 17606-2009 Determination of sulfur content in crude oil Energy dispersive X-ray fluorescence spectrometry |
detection range |
Determination of sulfur content in petroleum and petroleum products. |
Analyze the range |
S0.005~99.9%。 |
accuracy of analysis |
When the sample concentration is 100ppm< ±15ppm When the sample concentration is 1000ppm< ±5% When the sample concentration is 1%< ±1% |
detection limit |
15ppm |
Time |
100 seconds, 300 seconds, automatic, manual settings. |
Sample consumption |
4-10 milliliters |
external output |
USB2.0 |
Conditions of Use |
Environmental temperature: 5~+40 ℃, Relative humidity: ≤ 85%(30℃) Power supply: 220V ± 20V, 50Hz. |
Overall Power Consumption |
<150W |
Size |
Length x Width x Height: 430 x 390 x 130mm |
Weight |
12Kg |