Wavelength meter
Wavelength Meter
HighFinesse provides sensitive and compact wavelength meters with a wide spectral range for high-speed laser measurements. The optical unit is a photodiode array transmitted by a temperature controlled Fizeau based interferometer. High absolute accuracy is achieved through the use of solid-state, non moving optics. The optical unit and related electronic products are placed in a compact, thermal sleeve. Connect a computer or laptop through a high-speed USB 2.0 port, which allows for high data read rates. The analysis software displays all interferometer information and is compatible with Microsoft Windows.
Main features and advantages:
·Pulse and continuous wave lasers can be measured
·The most accurate instrument: Measurement resolution of ± 2 MHz, 500KHz
·High speed measurement up to 600 Hz
·Very wide spectral range: 192 nm -11 µ m
·Data control is achieved through simple API access
·(Virtual Instruments, C++, MathLab...)
Specific products:
1. High precision wavelength meter
Compact, reliable, and versatile
Plug and play
Customizable
Technical information:
measuring range |
WS5 |
WS6-600 |
WS6-200 |
WS7 |
Standard |
x |
x |
x |
x |
UV-I |
x |
x |
x |
x |
UV-II |
x |
x |
x |
x |
VisIR |
x |
x |
x |
|
IR-I |
x |
x |
x |
x |
IR-II |
x |
x |
x |
x |
IR-III |
x |
x |
||
Absolute accuracy1 |
||||
192-320 nm2 |
3 pm |
0.6 pm |
0.4 pm |
0.2 pm |
320-1100 nm |
3000 MHz |
600 MHz |
200 MHz |
60 MHz |
1100-2250 nm |
2000 MHz |
400 MHz |
150 MHz |
40 MHz |
2250-11000 nm |
2000 MHz |
200 MHz |
||
Quick Coupling Accuracy2 |
3000 MHz |
600 MHz |
600 MHz |
150 MHz |
Measurement rate |
||||
up to |
600 Hz |
600 Hz |
500 Hz |
500 Hz |
1) According to 3σ criteria |
2. Ultra precision series
Ultra high precision
Plug and play
Customizable
Technical information:
measuring range |
WSU2 |
WSU10 |
WSU30 |
|
Standard |
x |
x |
x |
|
UV-I |
x |
|||
IR-I |
x |
x |
||
Absolute accuracy1 |
||||
192-320 nm2 |
0.1 pm |
0.1 pm |
0.1 pm |
|
320-1100 nm |
2 MHz |
10 MHz |
30 MHz |
|
1100-2250 nm |
10 MHz |
20 MHz |
||
Quick Coupling Accuracy2 |
100 MHz |
100 MHz |
100 MHz |
|
Measurement rate |
||||
up to |
500 Hz |
500 Hz |
500 Hz |
|
1) According to 3σ criteria |