The micro moisture meter utilizes the principle of electrolytic detection, and its sensor is made of a cylindrical glass material and two parallel electrodes (usually made of platinum or rhodium metal wire, and the specific material needs to be selected according to the application), with a thin layer of phosphate H coated between the two electrodes3PO4The electrolytic current occurring between the two electrodes causes the water in the sample gas to decompose into H2And O2The electrode current is proportional to the moisture content in the sample gas, and the instrument processes the current signal in a series of ways to display and output data.It provides an absolute measurement of the trace water content in the sample gas.This principle can be used to directly measure the trace water content in all gases, includingIncluding Cl2、 HCl、 H2S、 H2SO4、 HBr、 SO2、 SF6、 CO2Wait for gases and all inert gases, except for a few gases that undergo chemical reactions with phosphoric acid.
It is possible to provide sensors with different material casings for different application conditions. The special sensor housing design allows the sample gas to flow through the sensor in a special transmission manner, ensuring rapid sensor response and minimal interference. These designs are suitable for very low ppmLevel measurement is very important.
The electrical connector between the sensor and analyzer is a waterproof sealed structure. Users can easily complete the regeneration of the probe within five minutes.
characteristic:
Made in Germany: High cost-effectiveness
Quick response and low gas consumption
Electrolytic method for direct detection of trace water in nitrogen gas, with a wide range of measurement and high accuracy
You can choose to include a rechargeable battery and a handle/arm
You can choose to include a card reader, which is availableSDCard records trace water content data (data can be output to)ExcelTable file)
Sensors are non consumable and renewable
midnight/Automatic range calibration function
Technical parameters:
Range:0- 10ppm/v、0- 100 ppm/v、0- 1000ppm/v、0- 2500 ppm/vMaintain the measuring range(7700ppm/v, used for sensor regeneration)
Accuracy: better than full scale±1 %
Sensitivity: Full scale±0.1 %
Operation: manual, remote control, automatic range switchingRS232
Response time:< 1In seconds,T50:< 8second
call the police:1Road, adjustable alarm concentration, dry junction relay.
Analog output:4-20mA (< 500 Ohm)
Display: LCD multifunctional display
Calibration: Zero point/Automatic range calibration
Power supply:80-230VAC; 50-60 HzOr rechargeable battery
power waste:8 W
Working temperature:+5 °Carrive+50 °C
Shell protection level: IP 20