Suitable for the power industry, power equipment manufacturing industry, etc., gas chromatography is used to determine the component content of dissolved gases in insulating oil, and can also be used for the determination of gas content in insulating oil.instrumentThe device adopts the principle of vacuum degassing - the automatic degassing method of variable diameter piston pump, which meets the technical requirements of the instrument in the national standard GB/T17623-1998 "Gas Chromatography Determination of Dissolved Gas Components in Insulation Oil".
1. The vacuum degree of the entire system shall not be less than 13.3Pa.
2. Degassing rate ≥ 95%, oil sample volume 15-20mL; When used for measuring gas content in oil, the inlet volume is 50mL.
3. The air volume for each wash is about 0.5mL, and it will automatically wash 6 times, 2.5mL ≤ total amount of air washing ≤ 3.5mL
4. The degassing process takes about 15 minutes to run.
5. The maximum vacuum degree of the vacuum pump is ≤ 6 × 10-2Pa; pumping rate 2 liters per second.
6. Power supply: AC 220V ± 10% 50Hz.