detailed description
The RV-300A and B-type hot deformation and Vicat softening point temperature measuring instruments are used to measure the hot deformation temperature and Vicat softening point temperature of thermoplastic materials such as plastics and rubber. Widely used in the production and scientific research of plastic raw materials and products. It adopts microcomputer control technology, automatic temperature and deformation measurement and control, and automatic calculation of test results. In addition, the machine has a compact structure, beautiful appearance, stable quality, and has the functions of eliminating oil smoke and odor pollution and cooling. It is a cost-effective model that is highly welcomed by users.
The RV-300A model is a desktop structure, while the RV-300B model is a floor standing structure.
The instrument complies with the requirements of ISO75, ISO306, GB1633, GB1634, GB8802 and other standards.
Main performance indicators:
1 Temperature control and measurement range: Room temperature -300 ℃
2 Heating rate: 120 ℃/h (12 ± 1 ℃/6 min)
50 ℃/h (5 ± 0.5 ℃/6 min)
3 Maximum temperature error: ±0.5℃
4 Maximum deformation error: 0.01mm
5 Number of sample racks: three
The RV-300E and F-type hot deformation and Vicat softening point temperature measuring instruments adopt microcomputer control technology and WINDOWS Chinese (English) interface to automatically measure and control temperature and deformation, draw real-time test curves, automatically obtain test results, and print test reports. This
The machine is currently the most advanced domestically produced model.
The RV-300E model is a desktop structure, while the RV-300F model is a floor standing structure.
The instrument complies with the requirements of ISO75, ISO306, GB1633, GB1634, GB8802 and other standards.
Main performance indicators:
1. Temperature control and measurement range: Room temperature -300 ℃
2. Heating rate: 120 ℃/h (12 ± 1 ℃/6 min)
50 ℃/h (5 ± 0.5 ℃/6 min)
3. Maximum temperature error: ±0.5℃
4. Maximum deformation error: 0.01mm
5. Number of sample racks: three
6. It has the functions of eliminating oil fume odor pollution and cooling