brand:Shanghai Jingxi Instrument Manufacturing Co., Ltd
summary:
SYD-267The manufacturing of the open flash point tester conforms to the industry standards of the People's Republic of ChinaSH/T 0318Technical Conditions for Open Flash Point Testers and Shanghai Enterprise StandardsQ/YXYY 23《SYD-267The requirements specified in the "Open Flash Point Tester for Petroleum Products" are applicable to the national standard of the People's Republic of ChinatreetopaccurateGB/T 267The method specified in the "Determination of Flash Point and Ignition Point of Petroleum Products (Open Cup Method)" is used to determine the flash point and ignition point of lubricating oils and dark petroleum products.
1、 Main technical features
1TheSYD-267The design of the open flash point tester fully complies withSH/T 0318、GB/ T 267Standard requirements.
2TheSYD-267The heating power of the open flash point tester is continuously and infinitely adjustable, and the power size is intuitively displayed by the ammeter. The temperature control method is advanced and reasonable.
3TheSYD-267The open flash point tester adopts a desktop and compact structural design, which is easy to operate and convenient to use.
SYD-267The biggest feature of the open flash point tester is its simple structure, continuous and infinitely adjustable heating power.
2、 Main technical parameters and indicators
1Working power supply:AC(220±10%)V,50Hz.
2Outer crucible ⑴ Material: High quality carbon structural steel, surface coated with black;
⑵ Inner diameter of upper opening: Φ100㎜±5㎜;
⑶ Height:50㎜±5㎜;
⑷ Bottom inner diameter: Φ56㎜±2Mm.
3Inner crucible ⑴ Material: High quality carbon structural steel, surface coated with black;
⑵ Inner diameter of upper opening: Φ64㎜±1㎜;
⑶ Height:47㎜±1㎜;
⑷ Bottom inner diameter:38㎜±1㎜;
⑸ Scale line: located at the edge of the upper mouth12Mm and18There is one engraved line at each millimeter.
4Gas conduit: nozzle diameter Φ0.8㎜~Φ1Mm, the smooth surface of the inner hole can adjust the flame3Mm~4mm.
5Heater: electric furnace heating, heating power1000WAdjustable.
6Temperature gauge:0~360)℃, division1℃, technical conditions meetGB/T 514The regulations.
7Power consumption of the whole machine: not greater than1100W.