Brand: Shanghai Jingxi Instrument Manufacturing Co., Ltd
This instrument is based on the industry standards of the People's Republic of ChinaJTG E20-2011In the "Test Regulations for Asphalt and Asphalt Mixtures in Highway Engineering"T0619-2011Asphalt kinematic viscosity test (capillary method), standard of the People's Republic of ChinaGB/T 1841-1980Designed and manufactured according to the requirements specified in the "Test Method for Viscosity of Polyolefin Resin Solutions", applicable toT0619-2011The requirements specified in the standard for determining the kinematic viscosity of asphalt samples at specified temperatures also apply toGB/T 1841-1980The standard requires the determination of the viscosity of decahydronaphthalene solution of polyethylene and polypropylene at a specified temperature.
1、 Main technical features
1. Adopting high-precision, digital display temperature controller and electric mixing device, the temperature control accuracy reaches ±0.1℃.
2. Adopting high temperature resistant, hard glass cylinder and insulated outer cylinder (referred to as double cylinder), with good insulation performance and clear observation of the sample.
3. Adopting a desktop and all-in-one design, the instrument has good overall integrity and is easy to use.
4. The maximum heating temperature of the constant temperature bath of this instrument can reach135℃.
2、 Main technical indicators and parameters
1. Working power supply:AC(220±10%)V、50Hz;
2. Heating power of bath solution: two levels,1000W(Auxiliary heating)+ 600W(Temperature controlled heating);
3. Bath liquid usage temperature: room temperature~135.0℃;
4. Oil bath temperature control accuracy: ±0.1℃;
5. Mercury thermometer: rod type, division value0.1℃; The temperature measurement range is(100~150)℃;
(150~200)℃ is optional;
6. Oil bath capacity:20L;
7. Sample quantity: Capillary suspension number:4Root;
8. Mixing motor: power6W; Speed1200r/min;
9. Environmental temperature:-10℃~+35℃;
10. Relative humidity:<85%;
11. Temperature sensor: Industrial platinum resistance, with a division number ofPt100;
12. Whole machine power consumption: not greater than1800W;
13. Capillary viscometer: A set of seven Kanfen counter current capillary viscometers, with models of200、300、350、400、450、500、600(approachRThe inner diameters of the pipes are1.02Mm1.26Mm1.48Mm1.88Mm2.20Mm3.10Mm4.00㎜).
SYD-265EPetroleum product kinematic viscosity tester(135℃Capillary method)