Brand: Shanghai Jingxi Instrument Manufacturing Co., Ltd
1Purpose and Scope of Application
SYD-261-The closed flash point tester for petroleum products is suitable for testing according toGB/T261-83、ASTMD93According to standard regulations,
Measure the lowest temperature of a mixture of steam and surrounding air formed when petroleum products are heated in a sealed oil cup and come into contact with a flame, causing a flash fire.
1The instrument should be placed on a flat and sturdy workbench.
2Rated voltage of the instrument220V±10%, frequency50Hz±1Hz.
3The studio power supply must have a good grounding device.
3、 Main technical specifications and parameters
1Thermal device:
voltage regulatorR-12 10A
Electric heating wire600W
Heating rate of oil testing 1-12 ℃/minute
2. Electric mixing device:
Constant speed motor45TCY
Transmission mode flexible shaft
Mixing blade size8×40(㎜)
stirring speed105 revolutions per minute
3. Standard oil cup:
inside diameter50.8㎜
depth56㎜
Oil testing capacity scale line34.2㎜
Oil testing capacity make an appointment70ml
4. Ignition device:
Ignition source: coal gas, butane gas, etc
Igniter aperture0.8㎜
5. Temperature gauge:
Internal standard or auxiliary mercury thermometers should comply withGB514-83。
scale-5-110 ℃, with a division of 1 ℃