brand | ZKP/Zhongkepu |
---|
Sec butanol is an important chemical raw material and intermediate, which can be used as a raw material for the production of methyl ethyl ketone, spices, dyes, lubricants, flotation agents, as well as solvents, emulsifiers, dye dispersants, dehydrators, paint removers, and industrial detergents. Tengzhou Zhongkepu adopts gas chromatographyQualitative analysis of impurities in industrial sec butanol products was carried out using chromatography, and the target components for gas chromatography analysis were determined. At the same time, R tx-200 capillary chromatography column was used to analyze the purity and impurity content of sec butanol. All impurities were well separated, and the accuracy and repeatability of quantitative analysis were good.
Instrument configuration and reagents
Product Name |
Configuration plan |
quantity |
Gas chromatograph |
GC-2020Equipped with (hydrogen flame detector+capillary injection system) |
1platform |
data processing |
Chromatography workstation |
1set |
Chromatographic column |
R tx - 200 ( 0. 32mm*60m*0. 5um ) |
1only |
External air source |
Nitrogen hydrogen air cylinder or generator |
1set |
computer |
Brand computer XP system |
1platform |
GC-2020 gas chromatographPerformance characteristics
1. Adopting320 x 240lattice5.7An inch large screen LCD Chinese display, real-time display of various temperatures and operating conditions, clear and intuitive content, truly realizing human-machine dialogue;
2Power on self-test, wide range self diagnostic function, can accurately determine the fault location and alarm;
3Six independent temperature controls (vaporization chamber and capillary vaporization chamber can be independently controlled), Five level program heating function;
4Overtemperature Protection function: If any circuit exceeds the set temperature, the instrument will automatically power off and sound an alarm;
5The vertical heating device of * makes sample vaporization more reliable, reduces the heat radiation generated in the vaporization chamber to the minimum, and ensures minimal temperature deviation inside the column box;
6Intelligent fuzzy control rear door opening system, automatically tracks temperature and dynamically adjusts The angle of the air door can achieve precise temperature control even near room temperature;
7Configure filling column head injection, glass lining injection, with diaphragm cleaning Functional capillary diversion/Non split injection device and capable of installing gas injectors;
8High precision dual stable gas path, capable of installing four types of detectors simultaneously;
9Optional electronic flow pressure display system with built-in workstation device.
Temperature control indicators
1.Temperature control range: room temperature 5℃~400℃increment0.1℃
2.Temperature control accuracy: superior to±0.01℃
Program heating: Five step interval constant temperature time0~999minincrement0.1minTemperature increment0.1℃
Heating rate:200℃The biggest one below40℃/min200℃Up to the biggest20℃/min