The portable online glove leak detector is a specialized leak detection device for detecting whether gloves have leaks. It is mainly used in sterile transfer chambers, negative pressure glove boxes, and sterile positive pressure/The online detection of gloves used for negative pressure isolators, sterile test isolators, etc. consists of high and low pressure inflation pumps, electrical and pneumatic control components, sensors, touch screens, glove box inflation docking units, outer boxes, and other parts.
No additional equipment is required for testing, no external compressed air source is needed, no gloves need to be disassembled, and the leak detection speed is fast, reducing the number of testing personnel and labor intensity; The testing unit adopts pressure resistant flexible connection pipelines, which can meet the leakage detection requirements of gloves at different heights; The button operation on the controller for start and stop, with a user-friendly design, greatly shortens the testing time and makes the testing operation faster and more convenient; Real time monitoring and printing of measurement data, three-level user password management, facilitating the subdivision of responsibilities and preventing misoperation.
Main technical parameters:
Working power supply:AV220 50HZ
Rated power:350W
Detection cycle:2-8 minutes per session
Test pressure:100Pa~5000 Pa
Measurement display accuracy:0.01Pa
Noise:≤50dB
Working mode: Continuous work
Number of test ports:1 piece
Test port size: elliptical, circular inner diameter (provided by the purchaser)
Weight:10Kg
Dimensions (length)X width; X height): 480mm × 230mm × 365mm (portable)
1. Preparation before operation
1.1. Check if the power supply is functioning properly.
1.2. Check if the instrument is intact and if the sealing ring is complete.
1.3. Check if the parameter settings are correct: Test pressure set at 2000Pa, qualified threshold set at 1500Pa, test time set30 seconds.
2. Instrument operation
2.1. Put the gloves on the glove tray corresponding to the glove leak detector, seal the gloves with the corresponding glove sealing ring, and pay attention to the tightnessThe sealing ring must be inserted into the guide groove.
2.2Connect the power cord, turn on the main power switch of the instrument, and enter the homepage work interface.
3. Maintenance and upkeep
3.1 Do not use strong acidic or corrosive solvents to wipe the glove leak detector during cleaning. Please use softer itemsclean,Stubborn dirt can be cleaned with neutral detergent, and dusty utensils should not be used for cleaning.
3.2 Inflatable sealing rings should not come into contact with sharp objects to prevent damage.
3.3 Regularly inspect and replace vulnerable parts such as inflatable seals.