Home>Products>Polymetrion9210 Silicon Meter Analyzer
Polymetrion9210 Silicon Meter Analyzer
● Multi channel operation (1-6) ● Extremely low operating cost, open reagent formula ● Small operation and maintenance volume ● Friendly interface, ea
Product details

Polymetron9210 silicon meter analyzerMeasurement principle:

The sample water enters the instrument through the sample water selection solenoid valve, and the flow rate of each sample water can be adjusted through the needle valve. Before the sample water enters the measuring pool
It has enough flow time to flush the entire waterway and overflow tank. Then, the sample water valve opens and the sample water enters the measuring tank. Once the measuring pool is flushed
After washing and filling with sample water, close the sample water valve and inject the reagents in sequence.
Reagent 1: Sulfuric acid+molybdate
Reagent 2: Oxalic acid
Reagent 3: Ammonium ferrous sulfate

Attention to the use of Polymetrion9210 silicon meter: For the delivery of reagents, micro piston pumps have now replaced traditional peristaltic pumps
The measuring cell is equipped with a heater and a magnetic stirrer to ensure thorough mixing and reaction of the reagents. Before the formation of phosphorus molybdenum blue moderator, the instrument
To measure the reference optical density, the absorption of light is calculated, and the concentration of phosphorus is calculated by comparing the calibration curve

Technical specifications of Polymetrion9210 silicon meter analyzer:

measuring range 0-1000ppb SiO2 0-5000ppb SiO2
reproducibility ± 0.5ppb or ± 2%, whichever is greater ± 2ppb or ± 2%, whichever is larger
detection limit ≤0.5ppb ≤2ppb
cycle time 10 minutes (per channel) 10 minutes (per channel)
sample requirements 5-50 ℃, 0.2-2 bar (optional 0.5-6 bar)
entrance 6mm outer diameter pipe PE/PTFE
emission 12 × 10mm inner diameter pipe PE/PTFE
power supply 100-240VAC,50/60Hz
power consumption About 80VA
calibration Automatic, adjustable
maintenance Replace reagents and calibration solutions every 45 days
display Equipped with backlight function, LCD display concentration, self diagnosis, alarm status, calibration constant, historical data, trend curve
Analog Output 6 pieces of 0/4-20mA (one per channel), maximum load of 650 Ohm
relay 8 of them
communication Optional RS485 300-9600 baud rate, maximum multi site 32, JBUS/MODBUS,Profibus
protection grade IP65/NEMA 4X material, disk mounted, polystyrene
weight Plate mounted: 10Kg, wall mounted cabinet: 10Kg

Polymetrion9210 Silicon Meter Analyzer Order Number:

Serial Number name order No
1 interface board 09200=A=5000
2 9210 silicon meter CPU board with display 09200=A=1000
3 9211 phosphate meter CPU board with display 09200=A=1011
4 power panel 09200=A=2000
5 RS485 board (MODBUS) 09125=A=0485/2485
6 92XX internal bus communication board 09200=A=5500
7 2-meter internal connecting cable/4m/7m 370=302=034/304/307
8 921X measuring board (0~1000ppb SiO2/0~5ppm-PO4) 0921X=A=1500
9 9210 silicon meter measuring board (0-5000ppb SiO2) 09210=A=1510
10 9211 Phosphorus Meter Measurement Board (0~50ppm-PO4) 09210=A=1520
11 Pipe 4 × 6 (customized domestically)
12 Sample water solenoid valve, NPS0,8(0,2~6bar) 689=118=008
13 O-ring (flow rate adjustment screw), 4 × 1mm
14 Sample water flow rate adjustment screw 09210=A=0105
15 Sample water level detector 09210=A=0250
16 O-ring (liquid level detector), 38 × 2mm
17 Sample water solenoid valve (overflow container) 689=118=024
18 921X connection plate (mixer/heater) 09210=A=1234
19 921X mixer 09210=A=0400
20 Magnetic stirring rod 226=003=013
21 heating element 09097=C=0560
22 Pt100 temperature probe
23 921X optical cable 09210=A=0500
24 921X optical mirror 09210=C=0340
25 921X mirror spacer 09210=C=0330
26 O-ring (optical mirror), 22 × 1mm
27 Square cover (photometer cell) 09210=C=0310
28 Reagent tube, 1,6 × 3,2mm 590=050=060
29 Insert the reagent tube, 0.8 × 1.6mm
30 Reagent pump 695=114=001
31 Calibration/flushing pump 695=004=004
32 921X2 liter reagent tank 495=020=001
33 921X reagent tube countersunk head 09210=A=0600
34 9210 silicon surface chemical reagent (excluding sulfuric acid) 09210=C=7000
35 9211 Phosphorus Surface Chemical Reagent (0-5ppm PO4, without sulfuric acid) 09211=C=7000
36 9211 Phosphorus Surface Chemical Reagent (0-50ppmPO4, without sulfuric acid) 09211=C=7001
37 9210 silicon surface agent (annual dosage) 09210=A=0507