Product Overview
The Picarro G2131-i isotope and gas concentration analyzer is capable of flexible measurements in various applications, ranging from atmospheric and marine scientific research to the source and authenticity of food and beverages. This analyzer is capable of measuring δ 13C in carbon dioxide with an accuracy of less than 0.1 ‰. This analyzer is capable of synchronously measuring the gas concentrations of carbon dioxide (CO2) and methane (CH4) with accuracies of 200 parts per billion (ppb) and 50 parts per billion (ppb), respectively. Measure water vapor (H2O) with a precision of parts per million (ppm), and calibrate and report CO2 and CH4 with dry gas mole fractions.
Product Features
• Measure δ 13C in CO2 with an accuracy of>0.1 ‰
• Paired with peripheral devices, it can measure the δ 13C of various samples
• Simultaneous measurement of CO2 and CH4 gas concentrations
• Measure water vapor and report dry gas mole fraction
Main technical parameters
Picarro G2131-iPerformance specifications |
|
CO2 inδ13CThe accuracy (1-σ,1 Hourly window,5 Minute average) |
Ensure accuracy< 0.1‰, inCO2 Concentration greater than380 ppmTypical accuracy of time< 0.25‰, inCO2 The concentration is200 ppmTypical accuracy of time< 0.05‰, inCO2 Concentration greater than1000 ppmWhen |
CO2 Under standard temperature and pressureδ13CThe maximum drift(24 Within an hour,1 The difference between the maximum values of hourly averages |
< 0.5‰ |
CO2 Accuracy of concentration(30 In seconds,1-σ) |
200 ppb (12C ) / 10 ppb (13C ) |
CH4 Accuracy of concentration(30 In seconds,1-σ) |
50 ppb + 0.05% (12C )reading |
H2OAccuracy of concentration(30 In seconds,1-σ) |
100 ppm |
CO2 dynamic range |
Ensure the accuracy range is380–2000 ppmThe scope of work is0.01–0.4%useA0314Small sample injection module( SSIM2)Peripheral devices can analyze 100% purityCO2 Sample, the device includes gas dilution function. useSSIM2The minimum volume of each sample injected during the test is10 µlpureCO2(0.45 µmolor20 µg CO2)Or an equal amount in the airCO2. |
CH4 dynamic range |
Ensure the accuracy range is0–500 ppmThe scope of work is0–1000 ppm |
H2Odynamic range |
Ensure the accuracy range is0–2.4%The scope of work is0–5% |
Instantaneous response |
per minuteCO2 The rate of change is300 ppmWhen,δ13CTypical deviation values< 0.1‰ |
Environmental temperature dependence |
ensure< ±0.06‰ / ℃, typical value< ±0.025‰ / ℃ |
Measurement interval |
about 2 Seconds (including regular ones)H2OandCH4 Measurement) |
rise time/Descent time( 10–90% / 90–10% ) |
The typical value is approximately 30 second |
Application precautions |
WhenH2O、CO2 andCH4 When the concentration is much higher than the normal environmental level, or when there are other organic compounds, ammonia, ethane, ethylene, or sulfur-containing compounds present, the measurement may be disturbed.H2OWhen the isotopic ratio changes significantly, it may affect the measurement results. Users should use prepared standard laboratory samples for validation. For more details about the experimental conditions, please contact us. |
Picarro G2131-iAnalyzer specifications |
|
measuring technique |
Cavity oscillation spectroscopy( CRDS )technology |
Measurement pool temperature control |
±0.005 ℃ |
Measurement pool pressure control |
±0.0002Atmospheric pressure |
Shock and vibration testing |
Satisfy military needsMIL-STD 810FThe standard for impact and vibration testing |
Sample temperature |
-10 to45 ℃ |
Sample pressure |
300 to 1000 Tuo(40 to 133 Kilopascals) |
Sample flow rate |
stay760 Underneath less than 50 Standard milliliters per minute(sccm)(Usually around)25 sccm)No need for filtering |
Sample humidity |
relative humidity( RH )less than99%, in40 ℃Under non condensing conditions, there is no need for drying |
Environmental temperature range |
10 to35 ℃(Run time)-10 to50 ℃(During storage) |
ambient humidity |
relative humidity( RH )less than99%Under non condensing conditions |
enclosure |
Pump (external), keyboard, mouseLCDMonitor (optional) |
data output |
RS-232EthernetUSBSimulation (optional)0–10Fu |
Air inlet connector |
¼inchSwagelok ® |
External dimensions |
Analyzer:17 Inches widex 7Inch heightx 17.5Inch length(43 x 18 x 45Centimeters), excluding 0.5 Inch foot external pump:5.6 Inches widex 6.4Inch heightx 11.9Inch length(14.3 x 16.3 x 30.3Centimeters) |
Installation form |
Workbench or 19 Inch rack mounted chassis |
weight |
60.4 Pound(27.4 Kg), including external pump |
Power requirements |
100–240Alternating current,47–63 Hz(Automatic detection), the total power at startup is less than 260 When the tile is in steady state, the analyzer is 125 Watt (analyzer), pump is 80 tile |