Nuclear Magnetic Resonance AnalyzerEquipped with independently developed high-performance spectrometers, high-power RF amplifiers, and gradient amplifiers, it has powerful functions. The product has a compact body and is a multifunctional desktopNuclear Magnetic Resonance AnalyzerProvide support for your laboratory component nuclear magnetic resonance analysis platform; This equipment can meet multiple testing requirements in fields such as petroleum exploration, with objective and accurate measurement results, high accuracy, good repeatability, stable instrument performance, and very high cost-effectiveness. In addition, the measurement method of nuclear magnetic resonance technology itself has many advantages: fast analysis speed, simple and convenient operation, no impact on the health of operators and the surrounding environment, no need for special sample preparation, non-destructive testing, no need for consumables such as chemical reagents, no pollution of samples, and green environmental protection.
This device is a powerful, high-precision, high accuracy, and high sensitivity oil core analysis instrument based on low field nuclear magnetic resonance technology. This instrument has many powerful functions, including FID, Hahn echo, IR sequence, CPMG sequence, IR-CPMG, PGSE-CPMG and other related measurements, which can meet the needs of pore, permeability and saturation testing in the field of petroleum rock cores.
Low field magnetic resonance is used to obtain liquidity information in rocks saturated with oil and water. It is specially designed for core samples in oil field reservoirs and can be used by oil producers and core analysis laboratories.
Main application areas:
Geological logging - rapid evaluation of oil and gas reservoirs
Logging - Basic Experimental Research on Nuclear Magnetic Logging
Experimental research on oil and gas development and indoor evaluation of low-permeability oil and gas reservoirs
Technical indicators:
1. Magnet type: permanent magnet;
2. Magnetic field strength: 0.3 ± 0.05T;
3. RF transmission power: peak to peak output greater than 300W;
4. Probe coil diameter: 25mm (standard), 125mm (optional);
Working conditions:
1. Power requirements: 220V, 50Hz;
2. Working temperature: 22-28 ° C;
3. Environmental humidity: 30~70%;
Size specifications (length, width, height):
1. Magnet cabinet: 826mm × 1120mm × 1326mm
2. Electronic control cabinet: 800mm × 600mm × 1170mm