Product Introduction
MK-1000Non destructive sealing testerAlso known as vacuum attenuation leak detector, it adopts non-destructive testing method, also known as vacuum attenuation method, and is professionally suitable for micro leak detection of ampoules, penicillin bottles, injection bottles, freeze-dried powder injection bottles, and pre filled packaging samples.
Product Features
Non destructive testing methods were used to detect micro leaks in finished packaging. After testing, the samples were undamaged and did not affect normal use, effectively reducing testing costs
Used for detecting small leaks and identifying samples with large leaks, the system automatically determines whether they are qualified or unqualified based on the leakage situation
The accuracy of vacuum attenuation method can reach 0.21cm (approximately 5 microns)
Wide application range, with corresponding test chambers available for different samples, allowing users to easily replace them
Automatic printing of test results at the end of the experiment, without the need for manual intervention, ensuring data accuracy and objectivity
10 inch color touch screen, user-friendly operation more convenient
Equipped with a micro printer, USB data interface, supporting PC software measurement and control operation, and exchanging units of mbar and Pa
Automatically save historical experiment records, locally query, and export to EXCEL format for saving
User level permission settings, meeting GMP requirements, testing record auditing, and tracking functions
Testing standards
This instrument complies with multiple national and international standards:
YY/T 0681.18-2020: Test Methods for Aseptic Medical Device Packaging - Part 18: Non destructive Inspection of Packaging Leakage by Vacuum Attenuation Method, ASTM F2338-13 Standard Test Method for Packaging Leakage - Vacuum Attenuation Method, USP1207 United States Pharmacopeia Standard
Test application
Basic applications | |
---|---|
Suitable for micro leak detection of ampoules, penicillin bottles, injection bottles, freeze-dried powder injection bottles, and pre filled packaging samples |
Technical indicators
index | parameter |
---|---|
Screen size | 10 inch touch screen |
Vacuum degree | Excellent vacuum as low as 10Pa |
Accuracy | 0.25 level |
Detecting aperture accuracy | 5 μ m (optional 1 μ m, 3 μ m) |
Test chamber | Size and type customized according to the special requirements of the sample |
Testing System | Dual sensor technology |
External dimensions | 470 mm (L) × 360 mm (B) × 300 mm (H) |
Power Supply | AC 220 V 50 Hz |