Each pipette has undergone factory testing and calibration in accordance with the requirements of ISO8655 before leaving the factory, and the claimed inaccuracy and precision are superior to the specified standards of ISO8655.
During the use of pipettes, deviations may occur due to various reasons, and regular calibration is an important means to ensure their accuracy.
Characteristics of single channel pipette
·Lightweight and ergonomically designed with digital windows, making the set range clear at a glance
·Wide range of measurement, from 0.1 μ l to 10ml
·Using attachment tools allows for convenient and efficient calibration and maintenance
·Accurate liquid separation, each pipette is calibrated according to EN/ISO8655 standards
·The lower half can be sterilized under high temperature and high pressure
·Adopting new materials
The multi-channel pipette also has the following innovative features
·The nozzle ejector can simultaneously eject multiple suction nozzles, which is efficient and labor-saving
·8 and 12 pipettes of various ranges are suitable for standard 96 well plates
·The liquid head can rotate 360 degrees, making it convenient for pipetting
·Each nozzle connection has an independent piston device, making maintenance and repair convenient
·Special nozzle connector design, easy to observe the sealing condition of the suction nozzle
technical parameter
Micro PetteSingle channel manually adjustable pipette
range |
increment |
Measurement volume (μ l) |
Maximum allowable system error |
Maximum allowable random error |
||
(Inaccuracy) |
(Inaccurate) |
|||||
% |
μl |
% |
μl |
|||
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2% |
0.05 |
1.25 |
3% |
0.0375 |
3% |
0.0375 |
||
0.25 |
12% |
0.03 |
6% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1% |
0.1 |
||
2 |
3% |
0.06 |
2% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2% |
0.1 |
2% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1% |
0.5 |
0.40% |
0.2 |
||
10 |
3% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3% |
0.6 |
1% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3% |
60 |
0.60% |
12 |
8Dao manual adjustable pipette
range |
increment |
Measurement volume (μ l) |
Maximum allowable system error (Inaccuracy) |
Maximum allowable random error (Inaccurate) |
||
% |
μl |
% |
μl |
|||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||
1 |
4% |
0.04 |
4% |
0.04 |
||
5-50μl |
0.5μl |
50 |
1% |
0.5 |
0.50% |
0.25 |
25 |
1.50% |
0.375 |
1% |
0.25 |
||
5 |
3% |
0.15 |
2% |
0.1 |
||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
150 |
1% |
1.5 |
0.50% |
0.75 |
||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
12Dao manual adjustable pipette
range |
increment |
Measurement volume (μ l) |
Maximum allowable system error (Inaccuracy) |
Maximum allowable random error (Inaccurate) |
||
% |
μl |
% |
μl |
|||
0.5-10μl |
0.1μl |
10 |
1.50% |
0.15 |
1.50% |
0.15 |
5 |
2.50% |
0.125 |
2.50% |
0.125 |
||
1 |
4% |
0.04 |
4% |
0.04 |
||
5-50μl |
0.5μl |
50 |
1% |
0.5 |
0.50% |
0.25 |
25 |
1.50% |
0.375 |
1% |
0.25 |
||
5 |
3% |
0.15 |
2% |
0.1 |
||
50-300μl |
5μl |
300 |
0.70% |
2.1 |
0.25% |
0.75 |
150 |
1% |
1.5 |
0.50% |
0.75 |
||
50 |
1.50% |
0.75 |
0.80% |
0.4 |
Manually fix the pipette
range |
increment |
Measurement volume (μ l) |
Maximum allowable system error (Inaccuracy) |
Maximum allowable random error (Inaccurate) |
||
% |
μl |
% |
μl |
|||
5μl |
- |
5μl |
1.30% |
0.065 |
1.20% |
0.06 |
10μl |
- |
10μl |
0.80% |
0.08 |
0.80% |
0.08 |
20μl |
- |
20μl |
0.60% |
0.12 |
0.50% |
0.1 |
25μl |
- |
25μl |
0.50% |
0.125 |
0.30% |
0.075 |
50μl |
- |
50μl |
0.50% |
0.25 |
0.30% |
0.15 |
100μl |
- |
100μl |
0.50% |
0.5 |
0.30% |
0.3 |
200μl |
- |
200μl |
0.40% |
0.8 |
0.20% |
0.4 |
250μl |
-- |
250μl |
0.40% |
1 |
0.20% |
0.5 |
500μl |
- |
500μl |
0.30% |
1.5 |
0.20% |
1 |
1000μl |
- |
1000μl |
0.30% |
3 |
0.20% |
2 |
2000μl |
- |
2000μl |
0.30% |
6 |
0.15% |
3 |
5000μl |
- |
5000μl |
0.30% |
15 |
0.15% |
7.5 |
|