Shanghai Precision Instrument Co., Ltd
Home>Products>MicroPette manual pipette (adjustable, fixed)
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    021-65729118
  • Address
    No. 1195 Pingliang Road, Shanghai
Contact Now
MicroPette manual pipette (adjustable, fixed)
MicroPette manual pipette (adjustable, fixed)
Product details

Each pipette has undergone factory testing and calibration in accordance with the requirements of ISO8655 before leaving the factory, and the claimed inaccuracy and precision are superior to the specified standards of ISO8655.

During the use of pipettes, deviations may occur due to various reasons, and regular calibration is an important means to ensure their accuracy.

Characteristics of single channel pipette

·Lightweight and ergonomically designed with digital windows, making the set range clear at a glance

·Wide range of measurement, from 0.1 μ l to 10ml

·Using attachment tools allows for convenient and efficient calibration and maintenance

·Accurate liquid separation, each pipette is calibrated according to EN/ISO8655 standards

·The lower half can be sterilized under high temperature and high pressure

·Adopting new materials

The multi-channel pipette also has the following innovative features

·The nozzle ejector can simultaneously eject multiple suction nozzles, which is efficient and labor-saving

·8 and 12 pipettes of various ranges are suitable for standard 96 well plates

·The liquid head can rotate 360 degrees, making it convenient for pipetting

·Each nozzle connection has an independent piston device, making maintenance and repair convenient

·Special nozzle connector design, easy to observe the sealing condition of the suction nozzle

technical parameter

Micro PetteSingle channel manually adjustable pipette

range

increment

Measurement volume (μ l)

Maximum allowable system error

Maximum allowable random error

(Inaccuracy)

(Inaccurate)

%

μl

%

μl

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2%

0.05



1.25

3%

0.0375

3%

0.0375



0.25

12%

0.03

6%

0.015

0.5-10μl

0.1μl

10

1%

0.1

0.80%

0.08



5

1.50%

0.075

1.50%

0.075



1

2.50%

0.025

1.50%

0.015

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08



10

1.20%

0.12

1%

0.1



2

3%

0.06

2%

0.04

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15



25

0.90%

0.225

0.60%

0.15



5

2%

0.1

2%

0.1

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15



50

1%

0.5

0.40%

0.2



10

3%

0.3

1.50%

0.15

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3



100

0.80%

0.8

0.30%

0.3



20

3%

0.6

1%

0.2

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3



100

0.80%

0.8

0.30%

0.3



50

1%

0.5

0.40%

0.2

100-1000μl

5μl

1000

0.60%

6

0.20%

2



500

0.70%

3.5

0.25%

1.25



100

2%

2

0.70%

0.7

200-1000μl

5μl

1000

0.60%

6

0.20%

2



500

0.70%

3.5

0.25%

1.25



200

0.90%

1.8

0.30%

0.6

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5



2500

0.60%

15

0.30%

7.5



1000

0.70%

7

0.30%

3

2-10ml

0.1ml

10ml

0.60%

60

0.20%

20



5ml

1.20%

60

0.30%

15



2ml

3%

60

0.60%

12

8Dao manual adjustable pipette

range

increment

Measurement volume (μ l)

Maximum allowable system error

(Inaccuracy)

Maximum allowable random error

(Inaccurate)

%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15



5

2.50%

0.125

2.50%

0.125



1

4%

0.04

4%

0.04

5-50μl

0.5μl

50

1%

0.5

0.50%

0.25



25

1.50%

0.375

1%

0.25



5

3%

0.15

2%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75



150

1%

1.5

0.50%

0.75



50

1.50%

0.75

0.80%

0.4

12Dao manual adjustable pipette

range

increment

Measurement volume (μ l)

Maximum allowable system error

(Inaccuracy)

Maximum allowable random error

(Inaccurate)

%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15



5

2.50%

0.125

2.50%

0.125



1

4%

0.04

4%

0.04

5-50μl

0.5μl

50

1%

0.5

0.50%

0.25



25

1.50%

0.375

1%

0.25



5

3%

0.15

2%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75



150

1%

1.5

0.50%

0.75



50

1.50%

0.75

0.80%

0.4

Manually fix the pipette

range

increment

Measurement volume (μ l)

Maximum allowable system error

(Inaccuracy)

Maximum allowable random error

(Inaccurate)

%

μl

%

μl

5μl

-

5μl

1.30%

0.065

1.20%

0.06

10μl

-

10μl

0.80%

0.08

0.80%

0.08

20μl

-

20μl

0.60%

0.12

0.50%

0.1

25μl

-

25μl

0.50%

0.125

0.30%

0.075

50μl

-

50μl

0.50%

0.25

0.30%

0.15

100μl

-

100μl

0.50%

0.5

0.30%

0.3

200μl

-

200μl

0.40%

0.8

0.20%

0.4

250μl

--

250μl

0.40%

1

0.20%

0.5

500μl

-

500μl

0.30%

1.5

0.20%

1

1000μl

-

1000μl

0.30%

3

0.20%

2

2000μl

-

2000μl

0.30%

6

0.15%

3

5000μl

-

5000μl

0.30%

15

0.15%

7.5

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!