purpose:
Direct measurement of environmental water samplesN2、O2andArDetection of concentration and stable isotopes; Can also be used forCO2、CH4、DMSMeasurement of dissolved gases with low molecular weight and other gases
Technical Parameter
1.1 Overall indicators:
Single gas(N2、O2、CO2)Coefficient of variation:0.1%
Gas ratio(N2/ArandO2/Ar)Coefficient of variation:0.03%
Sample required quantity:≤10ml
Sample analysis time:≤2min
Breathability:≥90%
1.2 Mass spectrometer:
Detector type: Faraday type detector
Unit resolution:<10%
Maximum work pressure:1×104MPa
Reproducibility of peak to peak ratio (under normal conditions)8hbelowN2andAr):0.5%
Can be networked through Ethernet
Detection limit:5×10-12mbar
Quality range:1-100u
Measurement speed:20ms—60s/amu
answerArSensitivity of:200MA/mbar
The impact on adjacent mass numbers:<10ppm
1.3 Peristaltic pump:
Speed:0.01to48rpm
Rotation mode: clockwise and counterclockwise dual modes
Flow range: Maximum:MPpattern:40ml/Minutes(1.8rise/Hour,4Millimeter tube), high-frequency mode:250ml/Minutes(15rise/Hour,8Millimeter tube
**:0.3rise/Divide(0.25Millimeter tube
Maximum pressure:MPThe mode is0.5 MPaIn high-frequency mode, it is0.3MPa
1.4 Vacuum extraction device
N2Injection rate:67rise/second
Front stage pump pumping speed@50Hz:0.5m3/h
extreme pressure:<1*10-7hpa/<7.5*10-8Torr
1.5 Thermostatic water bath controller:
Temperature stability:0.04°C
Maximum pressure:3.5psi(0.24bar)
Maximum pressure rate:2.9gpm(11.0 jpm)
Water capacity:28L
Temperature control range: ambient temperature+10°C—170°C
2Basic configuration
1) Main unit of membrane injection mass spectrometer:1cover
2) Peristaltic injection pump:1cover
3) Vacuum extraction device:1cover
4) Constant temperature water bath:2cover
5) Combustion furnace:1cover
6) Supporting accessories, etc.:1cover