Suzhou Bangxin Electronic Technology Co., Ltd
Home>Products>high performance liquid chromatograph
high performance liquid chromatograph
Used liquid chromatography instrument LC-20A (used gas chromatograph): infusion unit LC-20AD/20AT/20AB infusion method: parallel double plunger plunge
Product details

Used liquid chromatography instrument LC-20A (used gas chromatograph)

Infusion unit Prominence LC-20AD/20AT/20AB
LC-20AD
Infusion method: Parallel double plunger
Piston capacity: 10ul
*Large discharge pressure: 40MPa
Flow setting range: 0.0001mL/min-10.0000mL/min
Flow accuracy: within 1% or 0.5uL/min, whichever is greater (0.01mL/min-2mL/min)
Flow precision: below 0.3 (RSD: below 0.1%)
LC-20AT
Infusion method: Series double plunger
Piston capacity: main pump head 47ul, auxiliary pump head 23ul
*Large discharge pressure: 40MPa
Flow setting range: 0.001mL/min-10.000mL/min
Flow accuracy: 2% or 2uL/min, with the larger value within 0.01mL/min-5mL/min
Flow precision: below 0.3 (RSD: below 0.1%)
LC-20AB
Infusion method: Parallel double plunger (Type 2)
Piston capacity: 10uL
*Large discharge pressure: 40MPa
Flow setting range: 0.0001mL/min-10.0000mL/min
Flow accuracy: within the larger of 1% or 0.5uL/min (0.01mL/min-2mL/min)
Flow precision: below 0.3 (RSD: below 0.1%)
Gradient system
LC-20AB
Gradient method: high-pressure mixing
Number of mixed solvents: 2 liquids
Mixing concentration accuracy: below 0.5%
LC-20AD/20ATHigh voltage GE specifications
Gradient method: high-pressure mixing
Number of mixed solvents: 2 or 3 liquids
Mixed concentration accuracy: below 0.5% (LC-20AD), below 1% (LC-20AT)
LC-20AD/20ATLow voltage GE specifications
Gradient method: low-pressure mixing
Number of mixed solvents: * 4 more liquids
Mixed concentration accuracy: below 0.5% (LC-20AD), below 1% (LC-20AT)
UV Visible Detector
Prominence SPD-20A/20AV UV-VIS detector
Prominence SPD-M20A PDA detector
SPD-20A / SPD-20AV
Light source: SPD-20A D2 light, SPD-20AV D2 light, W light
Number of diode components: none
Wavelength range: SPD-20A 190nm-700nm, SPD-20AV 190nm-900nm
Spectral bandwidth • Gap width: 8nm
Wavelength accuracy: below 1nm
Wavelength precision: below 0.1nm
Noise: ± 0.25 × 10-5AU (under specified conditions)
Drift: 1 × 10-4AU/h (under specified conditions)
Linear: 2.5AU (ASTM specification)
Function: Dual wavelength detection at 190nm-370nm or above, proportional chromatogram output, wavelength scanning
Pool: Optical path length: 10mm, capacity: 12mL, pressure resistance: 12MPa
SPD-M20A
Light source: D2 lamp, W lamp
Number of diode components: 512
Wavelength range: 190nm-800nm
Spectral bandwidth • Gap width: 1.2nm (high-resolution mode) 8nm (high-sensitivity mode)
Wavelength accuracy: below 1nm
Wavelength precision: below 0.1nm
Noise: ± 0.3 × 10-5AU (under specified conditions)
Drift: 0.5 × 10-3AU/h (under specified conditions)
Linear: 2.0AU (ASTM specification)
Function: Contour line output, spectral retrieval, * large output
Pool: Optical path length: 10mm, capacity: 10mL, pressure resistance: 12MPa
Prominence SIL-20A/20AC automatic sampler
Injection method: full volume injection, variable injection volume
Injection volume setting range: 0.1ul -100ul (standard), 1ul -2000mL (optional)
Sample processing quantity:
175 (1mL sample bottle)
105 (1.5mL sample bottle) (SIL-20AC 70 (1.5mL sample bottle))
50 (4mL sample bottle)
192 (96 hole MTP 2 pieces)
768 (384 hole MTP 2 pieces)
In addition to the above specifications, an additional 10 1.5mL sample bottles can be used
Injection volume accuracy: less than 1%
Injection volume precision: RSD: Less than 0.3% (under specified conditions when injecting 10mL)
Cross contamination: less than 0.005% (naphthalene, chlorhexidine)
Repeated injection times: * 30/1 sample
Needle cleaning: set arbitrarily before and after injection
Sample cooler: SIL-20A without SIL-20AC module cooling and heating method, with moisture removal function at 4 ℃ -4
Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code