Dongguan Kunlun Industrial Technology Co., Ltd
Home>Products>Leakage tracing test box
Leakage tracing test box
The leakage tracing test box is IEC60112 Simulation test items specified in UL 746A, GB/T4207, GB4706.1, ASTM D 3638-92 and other standards
Product details

Leakage tracing test box

Product Description of Leakage Traceability Test Box


1Leakage tracing test boxIt is IEC60112 UL 746A, GB/T4207, GB4706.1, ASTM D 3638-92 and other standards
Standardized simulation test projects.

2. The leakage tracing test box is a device that applies a certain voltage between platinum electrodes of a specified size (2mm × 5mm) on the surface of solid insulation materials, and drips a specified volume of conductive liquid (0.1% NH 4 CL) at a fixed height (35mm) for 30 seconds. It is used to evaluate the leakage resistance of the solid insulation material surface under the combined action of electric field and humid or polluted media, and to determine its relative electrical trace index (CT1) and electrical trace resistance index (PT1).

3Suitable for research, production, and quality inspection departments of lighting equipment, low-voltage appliances, household appliances, machine tool appliances, motors, power tools, electronic instruments, electrical instruments, information technology equipment, as well as insulation materials, engineering plastics, electrical connectors, and accessories industries.
Technical parameters of leakage trace testing box:
1. Electrode Material: Electrode Head - Platinum Metal (Platinum)
2. Electrode pressure: 1.0 ± 0.05N
3. Electrode distance: 4.0mm ± 0.01mm, angle 60 °± 5
4. Electrode voltage: adjustable from 100~600V (48~60HZ), voltage drop less than 10% when short-circuit current is 1.0 ± 0.1A
5. Droplet height: adjustable from 30~40mm
6. Drop size: 44~55 drops/1cm3 adjustable
7. Dropping time: 30s ± 5s,
8PTI testerDropping times: adjustable from 0 to 9999 times
9. Leakage judgment: If the circuit current is greater than 0.5A and maintained for 2 seconds, the relay will act to cut off the current, indicating that the sample is not qualified
10. Experimental process: Automatic control of the experimental program, independent ventilation
11. PTI tester reference standard: IEC60112 、 GB/T4207、GB4706.1
12. Studio volume: 0.5 (can be customized to 0.75 m3 or 1m3 according to different standards)