Main functions:
l Highly integrated controller. Powerful data recording and analysis software that can control light intensity.
l Automatically collect data. Automatically calculate photosynthetic rate or respiration rate.
l Equipped with a controllable light source, the desired light intensity can be set by oneself, and photosynthesis can be automatically performed-Measurement of light intensity response curve and automatic calculation of photosynthetic quantum efficiency.
l Can be withHANSATECHThe company'sFMS 2Simultaneous measurement of photosynthetic oxygen release rate and changes in chlorophyll fluorescence using a fluorescence analyzer
Origin and Manufacturer: UKHansatech
1. Sample Room:7.5 ml
2. Sample area:10 cm2
3. Measurement range:0~40% O2
4. Oxygen resolution:20At ℃10×10-6μ mol· ml
5. controlmakeDevice: Computer controller and integrated magnetic stirrer, capable of controlling light intensity and stirrer speed, automatically collecting data(0.2~10order/Seconds),RS232output
6. Software Function: Powerful data recording and analysis software. Automatically record the dynamic changes in oxygen signals and calculate respiratory rate and photosynthetic rate. Automatic control of light intensity for illumination-Measurement of photosynthetic response curve and automatic calculation of photosynthetic quantum efficiency
7. Controllable light source: equipped with a red light source, with a light intensity of up to750 μ mol· m-2·s-1, can set the required light intensity by oneself
8. Temperature control: The sample reaction chamber has a circulating water bath interlayer, which can be connected to an external circulating water bath for temperature control
9. PARCalculation:Light intensity measurement range:0~2000 μ mol· m-2·s-1
10. Electrode output:21% O2 When is1 μA;10%~90%Range response time<5second
11. Residual current: <0.02 μA
12. Polarization voltage:700 mA
13. Working voltage:100~260V
14. Controller volume:250×126×60 mm;
15. Weight:650g