-Instrument Introduction
The LRS-2/3 laser Raman spectrometer is a new product designed by our company. It won the second bid for World Bank loan development projects in 1998 and 2000 due to its simple structure, easy adjustment and measurement, high sensitivity, good stability, and low price. Suitable as a measuring instrument for Raman spectroscopy and fluorescence spectroscopy in research institutes, physics laboratories, and chemistry laboratories of higher education institutions.
-Product Features
·computer control
·Can automatically record Raman and fluorescence spectra
·High resolution, low stray light monochrome system
·High sensitivity, low-noise single photon counter for receiving system
·High power semiconductor laser as excitation light source
·Equipped with a stable and high-precision external light path system
·Equipped with multiple accessories, suitable for analyzing liquid and solid samples
·Equipped with a notch filter for reducing Rayleigh scattering
-TECHNICAL INDEX
number |
name |
specifications |
1 |
monochromator |
D/F=1/5.5, raster1200L/mm, slit width0-2mmContinuously adjustable, slit accuracy0.01mm |
2 |
Semiconductor laser |
Output wavelength:532nmOutput power ≥ 100mW |
3 |
External light path system |
Including laser focusing optical system and Raman signal collection optical system. Optical components such as polarizer, analyzer, waveplate, Raman filter, etc. can be placed in the optical path |
4 |
Sample Holder |
Five dimensional adjustable vertical sample holder, horizontal liquid sample holder, transparent solid sample holder, oblique incidence solid sample holder, backscatter sample holder |
5 |
Single photon counting system |
Photon counting type photomultiplier tube, single photon signal discrimination circuit system |
-Basic configuration and parameters
number |
name |
parameter |
1 |
focal length |
f=302.5mm |
2 |
relative aperture |
D/F=1/5.5 |
3 |
Scan range |
0nm-1200nm |
4 |
linear dispersion |
2.7nm/mm |
5 |
resolution |
≤0.1nm(500nm) |
6 |
slit |
width0~2mm continuously adjustable, indication accuracy of 0.01mm, seam height up to 20mm |
7 |
modal |
Near TEM00 |
8 |
Wavelength accuracy |
≤0.1nm |
9 |
Wavelength Repeatability |
≤0.2nm |
10 |
stray |
≤10-3 |
11 |
External dimensions |
700mm×500mm×450mm |
12 |
weight |
70kg |