Dimensions: 180mm × 110mm × 40mm
(1) Intelligent host with all functions (LPE-1601)
Multiple additional measurement functions, multiple laser wavelength automatic compensation calibration, automatic range change, automatic zero compensation, normalization with standard values, additional attenuation coefficient compensation, and other intelligent measurement and processing, display functions, and data processing options.LPE-1601Ready and able to pass throughRS232/USBInterface connectionPCOutput measurement data to other computers and other devices; And can utilize specially designedPCThe software is used to expand measurement, display, and information processing functions.
(2) Main unit of laser power energy meter (LPE-1601-WJ)
LPE-1601-WJ series laser power/Energy meterIt is a laser radiation intensity measurement instrument that uses microprocessor technology; By host(LPE-1601),Composed of a T-shaped thermopile laser measurement detector selected by the user according to the measurement parameters.LPE-1601-WJlaser power/Energy meters are suitable for measuring the power of continuous lasers and the energy of pulsed lasers.
(3) Host for measuring continuous laser power (LPE-1601-W)
LPE-1601-W series laser power meterIt is a laser radiation intensity measurement instrument that uses microprocessor technology; By host(LPE-1601),Composed of a T-shaped thermopile laser measurement detector selected by the user according to the measurement parameters.LPE-1601-WLaser power meters are suitable for measuring the power of continuous lasers and the average power of high repetition rate pulsed lasers.
(4) Host for measuring pulse laser energy (LPE-1601-J)
LPE-1601-J laser energy meterIt is a laser radiation intensity measurement instrument that uses microprocessor technology; By hostLPE-1601,Composed of T-shaped thermopile or P-type pyroelectric detector laser measurement detector selected by the user according to the measurement parameters.LPE-1601-JLaser energy meters are suitable for measuring the energy of pulsed lasers.
T-shaped probe
model |
T-15 |
T-25 |
T-35 |
T-50 |
Sensor type |
Calorimetry |
Calorimetry |
Calorimetry |
Calorimetry |
Sensitive surface shape |
plane |
plane |
plane |
plane |
Spectral response range(mm) |
0.19-11 |
0.19-11 |
0.19-11 |
0.19-11 |
mm Spectral response uniformity(%) |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
Sensitive diameter(mm) |
15 |
25 |
35 |
50 |
sensitivity |
>100mV/W |
>60mV/W |
>20mV/W |
>20mV/W |
response time(ms) |
<1000 |
<2000 |
<10000 |
<10000 |
10nsPulse power density threshold(MW/cm2) |
100 |
100 |
100 |
100 |
The maximum power or energy |
2W/J |
20W/J |
50W/50J |
200W/50J |
Detector dimensions |
|
|
|
|
Price (RMB) |
|
|
|
|
P-type probe
model |
P-15 |
P-20/ P-25 |
P-35 |
P-50 |
Sensor type |
Pyroelectric |
Pyroelectric |
Pyroelectric |
Pyroelectric |
Sensitive surface shape |
plane |
plane |
plane |
plane |
Spectral response range(mm) |
0.19-11 |
0.19-11 |
0.19-11 |
0.19-11 |
mm Spectral response uniformity(%) |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
0.19-2.5 2.7 |
Sensitive diameter(mm) |
15 |
20/25 |
35 |
50 |
sensitivity |
>5mV/ mJ |
>5mV/mJ |
>2mV/ mJ |
>1mv/mJ |
response time(ms) |
<1 |
<5 |
<5 |
<10 |
10nsPulse power density threshold(MW/cm2) |
100 |
100 |
100 |
100 |
The maximum power or energy |
2J |
20J |
20J |
20J |
Detector dimensions |
|
|
|
|
Price (RMB) |
|
|
|
|