PCD-430A sensor interface amplifier
-----------------------------------------------------------------------
Summary
Just go throughUSBA measuring device that can be connected to a computer through an interface for measurement.
Through stack type connections, it can be easily extended to16A channel.
feature
Input strain and voltage can be selected for each channel
● Easy sensor connection based on abundant input adapters (1 input adapter is required for measurement)
The USB interface is used to connect to a personal computer
Possible measurement with 1 unit of 4 channels and a maximum of 4 units of 16 channels (allowing for a combination of PCD-400A)
● Conduct simultaneous sampling and investigation of 4 units using connector sets sold separately
Dynamic Data Set Recording Software (DCS-100A) Standard Accessory for Windows Vista, Windows 7, Windows 8, 8.1 (Japanese/English, 32/64 bits)
● TEDS Corresponding (UI-10A, UI-11A)
● Input adapter
Input adapter UI-10A for strain gauge sensors (corresponding to TEDS)
Strain gauge input adapter UI-11A (corresponding to TEDS)
Input adapter UI-15A with operating rod strain gauge
Input adapter UI-16A for single touch locking strain gauges
One touch input adapter UI-55A
Voltage input adapter UI-30A
● Small and lightweight
● Data analysis can be performed by selecting analysis software (DAS-200A, NI DIAdem)
Product Overview
Comprehensive specifications
| Number of measurement channels | 4 |
|---|---|
| Synchronize operation | Up to a maximum of 4 units with 16 channels |
| sampling | Sampling frequency: maximum 10kHz (when set to 10kHz, 4 sets of 16 channels can be measured synchronously) |
| AD converter resolution | 24 digits |
| Maintaining the set value | The range and balance adjustment values are saved in the internal non-volatile memory |
| TEDS function |
Read information from TEDS sensors (input adapter: only UI-10A, UI-11A is valid) There is a function to write channel names (valid when the manufacturer's ID is Gonghe Electric) |
| interface | USB 2.0 (High Speed specification, can also work with USB 3.0) |
| Temperature and humidity range for use | 0-40 ℃, 20-85% RH (no condensation) |
| Vibration resistance | ± 29.42 m/s2 (3G), 5-200Hz (12 cycles per axis, 10 minutes per cycle) |
| power supply |
DC11~16V Joint type: RM12BRD-4PH (Hirose Electric Co., Ltd.) |
| current consumption | Below 0.9A (DC12V) |
| Dimensions of appearance | 210 (W) × 35 (H) × 157.5 (D) mm (excluding protruding parts) |
| weight | About 750g |
Specifications
Applicable specifications
| EMC Directive | EN61326-1 (Class A) |
|---|---|
| RoHS Directive | EN50581 |
Strain measurement mode
| measurement object | Strain gauges, strain gauge sensors |
|---|---|
| input format | Balanced differential input |
| Suitable for strain gauge resistors | 1/4 bridge method (2-wire, 3-wire) 120 Ω Half bridge type, full bridge type 120 Ω~1000 Ω |
| Bridge power supply | AC 2Vrms |
| strain rate | 2.00 Fixed |
| Balance adjustment range | Resistance: ± 2% (± 10000 × 10-6 strain) or less Capacitor: 5000pF or below |
| Balance adjustment method | Resistance: Pure electronic automatic balancing method Capacitor: CST mode (automatic tracking) |
| nonlinear | ± 0.1% FS or less |
| measurement range | 200,500,1000,2000,5000,10000,20000×10-6The 7 segments of the dependent variable Accuracy: ± 0.5% FS or below |
| Response frequency range | DC~200Hz, Deviation within ± 10% or less |
| low-pass filter | Communication characteristics: 2nd order Butterworth type Cut off frequency: 10,30100Hz and 4 segments of FLAT Amplitude ratio at the cut-off point: -3 ± 1dB Attenuation characteristics: -12 ± 1 dB/oct. |
| stability | temperature Zero point: ± 0.2 × 10-6Strain/℃ Sensitivity: ± 0.05%/℃ time Zero point: ± 0.5 × 10-6Dependent variable/8h Sensitivity: ± 0.15%/8h |
| Insulation pressure resistance | Input - Between Chassis AC250V, 1 minute |
Voltage measurement mode
| measurement object | voltage |
|---|---|
| input format | Unbalanced input |
| measurement range | 6 sections of 1, 2, 5, 10, 20, and 50 V 精度: ±0.2%FS 以内 |
| Response frequency range | DC ~ 1kHz, Deviation -3 to+1dB or less |
| high-pass filter | Cut off frequency: 0.2Hz, 2 segments of OFF |
| low-pass filter | Communication characteristics: 2nd order Butterworth type Cut off frequency: 10, 30, 100, 300 Hz and 5 segments of FLAT Amplitude ratio at the cut-off point: -3 ± 1dB Attenuation characteristics: -12 ± 1 dB/oct. |
| stability | temperature Zero point: ± 0.008% FS/℃ Sensitivity: ± 0.02%/℃ time Zero point: ± 0.03% FS/8h Sensitivity: ± 0.1%/8h |
Standard Accessories
| Standard Accessories | USB cable N-38 (1m) Grounding wire P-72 (5m) DVD (Dynamic Data Control Software DCS-100A) |
|---|
Choose accessories
| Choose accessories | AC adapter UIA345-12 USB cable N-39 (2m) DC power cable P-76 (DC 11-16V, 2m) Connecting cable N-97 (10cm) Connection connector device (ST-1A) Connection board (CN-20) Input adapter UI-10A for strain gauge sensors (compatible with TEDS) Strain gauge input adapter UI-11A (compatible with TEDS) Input adapter UI-15A with operating rod strain gauge Input adapter UI-16A for single touch locking strain gauges One touch input adapter UI-55A Voltage input adapter UI-30A |
|---|